Technical data
5
Settings
............................
5.8 – 5.8.1
5.8 Data interface
........................................................................................................................................
The D190 remission densitometer features a serial RS232 interface with adjustable baud rate.
The following settings are available: No handshake, Software handshake (Xon/Xoff) or
Hardware handshake.
........................................................................................................................................
........................................................................................................................................
The Data interface function includes:
• Baud rate setting (110, 150, 300, 600, 1200, 2400, 4800, 9600 or 19200).
• Handshake setting: Xon/Xoff, Hardware or No.
........................................................................................................................................
Procedure:
• Select the menu line Data interface by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
• The menu for the configuration of the interface is shown on the display.
5.8.1 Baud rate setting
Procedure:
• Select the menu line Baud rate by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
• The entry field for the baud rate selection is shown in reverse video on the display.
• Enter the desired baud rate by turning the <Control ball> and terminate the entry by pressing the
<Measurement key>.
• Select the menu line Return by turning the <Control ball> and press the <Measurement key>
several times to return to the main menu.