User's Manual

Entry of Reference Values... By Numeric Entry
> Whiteness & Yellowness reference values can not be entered via the keyboard.
1) Select desired Illum/Obs (see
Sec. 3.3).
2) Press
[FUNCTION]
to select
desired measurement space.
> If is not displayed,
momentarily press
[DIF]
to
activate Difference mode.
3) Hold down
[DIF] until
"SET
REF VALUES - ENTER or
MEASURE" is displayed.
The unit will first display the
previous entered reference
values for the last selected
reference location.
The reference location is
momentarily displayed.
Then "EXIT" is displayed.
4) Select the reference location.
If the cursor is not blinking
next to the references
location, press [DIF] until it
does.
Hold down [DIF] and press
[t] to decrement thru the
reference locations or press
[s] to increment.
FUNCTION ILLUM
DIF
CAL
X + 4 . 1 2
Y 2 . 8 7
Z + 2 . 4 1
C
2
X Y Z
+
PRESS
FUNCTION ILLUM
DIF
CAL
SET REF VALUES
ENTER or MEASURE
XYZ
Z 2 . 6 9
C
2
Y 4 . 4 5
X 6 . 1 6
Z 2 . 6 9
Y 4 . 4 5
X 6 . 1 6
Z 2 . 6 9
Y 4 . 4 5
X 6 . 1 6
EXIT
REF 03
HOLD DOWN
FUNCTION ILLUM
DIF
CAL
Z 2 . 6 9
Y 4 . 4 5
X 6 . 1 6
CURSOR
+
REFERENCE SPACE
REF 03
DECREMENT
INCREMENT
+
16