Instruction manual

SECTION THREE
3-2
Page 2 Configuration Options
IO preset
This options provides several QC system selections. All the necessary
RS-232 interface parameters are automatically set when a predefined I/O
preset is selected. A “CUSTOM” preset is available which allows you
manually set all available I/O options. Below is a chart that contains all
available I/O presets and their parameter settings.
NOTE: Certain parameters are mandatory for the I/ O preset and cannot be
changed. They are labeled with an “L” for locked. Other parameters are not
considered mandatory and are labeled with a “U” for unlocked. Finally,
parameters that remain unchanged from previous settings—and can be
changed—are shown as an asterisk (*).
I/O Preset RCI PIN5 XON DPT COMP ALF DEL AXMT BAUD
REPORT Off * * On (L) On (L) On (U) Off (L) Off (U) *
SPRD-SHT * * * * Off (L) * * On (L) *
k:TNetA Off (L) Off (L) Off (L) Off (L) On (L) On (L) On (L) On (L) 300(L)
k:TNetXT Off (L) CTS(L) Off (L) Off (L) On (L) On (L) Off (L) On (L) *
K:C.A.P. Off (L) CTS(U) Off (L) Off (U) On (U) On (U) Off (L) On (L) 9600(L)
k:SYS25-75 Off (L) CTS(L) Off (L) Off (L) On (L) On (L) Off (L) On (L) *
f:TECOM Off (L) Off (L) Off (U) Off (U) Off (U) Off (U) Off (U) On (U) 1200(U)
c:I/O#1 On (L) Off (L) Off (L) Off (U) On (U) On (U) On (U) On (L) 9600(L)
c:I/O#2 On (U) Off (U) Off (U) Off (U) On (U) On (U) On (U) On (U) 9600(U)
MITSY#1 Off (U) Off (U) Off (U) Off (L) Off(L) Off(L) Off (U) On (U) 300(L)
MITSY#2 Off (L) Off (L) On (U) Off (L) Off(L) Off(L) Off (L) On (L) 300(U)
NORITSU Off (L) CTS(U) Off (L) Off (L) On (L) On (L) Off (L) On (L) 2400(U)
n:QSSnet Off (U) Off (U) Off (U) Off (U) Off (U) Off (U) Off (U) On (U) 2400(U)
n:Micro * * * * * * * * *
kn:Net TTL (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) Off (U) 2400(U)
a:LC2000 Off (U) Off (U) Off (U) Off (U) On (U) On (U) On (U) On (U) 1200(U)
CUSTOM # # # # # # # # #
Windense On (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) On (U) 9600 (U)
Kodatel On (U) Off (U) Off (U) Off (U) On (U) Off (U) Off (U) On (U) 9600 (U)
Sienna On (U) Off (U) Off (U) On (U) On (U) On (U) Off (U) On (U) 4800 (U)
k: (Kodak), f: (Fuji), c: (Copal), kn: (Konica), n: (Noritsu), a: (Agfa)
Page 2a Configuration Options
RCI (Remote Control Interface)
Controls the ability of the instrument to be operated remotely. Available
settings are: OFF, ON, TTL.
OFF - sets the instrument into a receive mode at which time the instrument
receives data.
ON - enables data to be received via RCI command.
TTL (Time Tagged Log) - causes all data received into the serial port to be
placed into the log buffer with a time stamp.
PIN5
Determines the status of the handshaking input on Pin 5 of the RS-232 I/O port.
Pin 5 may be interpreted as BUSY, CTS (clear-to-send), or OFF (ignored).