Specifications

Narda 19NBM-580
3�3 Connecting the sensor
3�3 Connecting the sensor
WARNING
Sensor is not operating properly
Possiblypresenthighradiationvaluescannotbe
recognizedwhenasensorisdefective.
Check sensors for proper operation with a
signal source before using this measuring
instrument.Thisisespeciallyimportantfor
thermocouples because the sensors can be af-
fectedbyvariousmechanicalandenvironmen-
talstressors.Nardaoffersportablesourcesto
accomplishthisimportantstep(seeAccesso-
ries on page 58).
BeforebeginninganyRFradiationmeasure-
ment,alwaysadviseyourselfofthefrequencies
andeldstrengthsthatyoucouldexpectto
encounter
Therearemanydifferentprobesavailablefordifferent
applicationsfortheNBM-580.Youcanndmoreinforma-
tionabouttheordernumbersandspecicationsofthe
probesunderOrderinginformationonpage56,aswell
asinthedatasheetsoftheNBM-580andprobes.These
documentscanalsobedownloadedfromtheNardaweb
site on the Internet at http://www.narda-sts.com
NOTICE:
Wrong handling of the probe
Damageoftheprobehead
Alwaysholdtheprobeatthemetalplugend.