Instruction manual
. .
3 Display 28
Figure 3-9: Data file of weights hung from a transducer and the units conversion dialog window that sets volts to grams.
Once recording is complete, proceed to the 2 Cursor mode in the Main window. (See the
Cursors section of this chapter.) The Units Conversion dialog window cannot be entered
without being in 2 Cursor mode. Position Cursor 1 over one of the known values, and Cursor 2
over the other known value.
Figure 3-10: The Units Conversion dialog window showing the 2-point calibration boxes.
Select the Units menu from the Channel menu or by clicking on the Value display area on the
channel bar. Select the Simple Units Option. The Units dialog window will appear. The typical
conversion of the output of a transducer is a 2-point calibration, which is the default conversion
method (Figure 3-10 on page 28). Below that menu is an area where the values for the positions
of the cursors are listed. The values on the left are the voltage values at the positions of