Instruction manual
SI-MT/SI-MKB/SI-HTB
WORLD PRECISION IN STRU MENTS
37
CH 1 Setpoint
37.0ºC
Fig. 39 Press the Up and Down buttons to adjust the Channel 1 Setpoint.
2. Press the Up or Down button to adjust the setpoint. The maximum setpoint allowed
is 45ºC.
3. Press the Display button to save the confi guration and return to the normal display.
Setting Alarms
Both Channel 1 and Channel 2 have high and low alarm values. By default, the low
alarms are set at 36ºF, the high alarms are set at 38ºF and the alarms are disabled.
1. Press the Setup button:
• Twice to display the Channel 1 High Alarm
• Three times to display the Channel 1 Low Alarm
• Five times to display the Channel 2 High Alarm
• Six times to display the Channel 2 Low Alarm
The alarm setting displays.
CH 1 High Alarm
38.0ºC
Fig. 40 Press the Up and Down buttons to adjust the alarm setting.
2. Press the Up or Down button to adjust the alarm setting.
3. Press the Display button to save the confi guration and return to the normal display.
Changing the Backlight Level for the Display
By default the backlight level is set at 4. To make the display brighter, increase the level
up to a maximum of 8. To dim the display, choose a lower level.
1. Press the Setup button until “Backlight Level” appears on the screen.
Backlight Level
Min=1 4 Max=8
Fig. 41 Press the Up or Down buttons to adjust the backlight level.
2. Press the Up or Down button to adjust the backlight level.
3. Press the Display button to save the confi guration and return to the normal display.