Instruction manual

SI-MT/SI-MKB/SI-HTB
WORLD PRECISION IN STRU MENTS
31
Adjusting the CCD Offset
CAUTION: The CCD offset should not be modifi ed from the factory default of
6.8mm unless you have a reason to place the zero order laser spot at a different
location other than the cross hair mark on the cuvette.
1. Press the Setup button until "CCD Offset" appears. This is the distance between the
crosshair marking on the cuvette and the fi rst pixel of the CCD camera.
CCD Offset
6.8 mm
Fig. 27 Press the Up or Down button to adjust the offset.
2. Press the Up or Down button to adjust the offset.
3. Press the Display button to save the confi guration and return to the normal display.
Setting the Tissue Distance
CAUTION: The default value for the tissue distance is 36.0mm. This value should
not be modifi ed.
1. Press the Setup button until " Tissue Distance" displays. This is the distance between
the muscle tissue sample and the CCD camera (D).
Tissue Distance
36.0 mm
Fig. 28 Press the Up or Down button to set the tissue distance.
2. Press the Up or Down button to adjust the distance.
3. Press the Display button to save the confi guration and return to the normal display.
Adjusting the CCD Sensitivity
1. Press the Setup button until “ CCD Sensitivity" displays. The camera's sensitivity can
range from 1 (minimum) to 10 (maximum). By default the sensitivity is set to the mini-
mum (1). To modify the sensitivity, press the Setup button.
CCD Sensitivity
Min=1 1 Max=10
Fig. 29 Press the Up or Down button to adjust the camera sensitivity.
2. Press the Up or Down button to adjust the sensitivity.
3. Press the Display button to save the confi guration and return to the normal display.