Instruction manual
SI-MT/SI-MKB/SI-HTB
WORLD PRECISION IN STRU MENTS
27
8. If you intend to use multiple amplifi cation factors, cross-check your calibration. For
example, if a 0.5g mass was used to calibrate the SI-BAM21-LCB as close as possible
to 10.0V at a gain of X10, then at a gain of X1, the monitor should display an output
very close to 1.0V for the same 0.5g calibration mass.
Making Measurements
After the SI-BAM21-LCB has been calibrated, measurements may be taken.
1. Turn the SI-BAM21-LCB Power switch on (I).
NOTE: Allow the system to stabilize for about 30 minutes. This allows all the compo-
nents to reach thermal equilibrium, minimizing measurement changes due to thermal
variations.
2. Turn on the data acquisition system.
3. Press the Zero button to set the baseline value for the measurements.
NOTE: When the Zero button is pressed, the zeroing LED illuminates to indicate that
it is functioning properly.
4. Measurements may be taken.
Setting System Gain Factor
The SI-BAM21-LCB gain multiplier setting is selected with an internal jumper that is
confi gured at the factory for use with the muscle tester system of your choice (SI-MT,
SI-MKB, SI-HTB). The X1 setting (SI-MT/SI-HTB) allows for 1X, 2X, 5X and 10X gains.
The X10 setting (SI-MKB) allows for 10X, 20X, 50X and 100X gains that may be needed
when recording passive tension or small muscle contractions.
1. Turn off the Signal Conditioning Amplifi er System and unplug it from the power out-
let.
2. Remove the two screws on the face of the SI-BAM21-LCB module.
3. Gently slide the module out of the Signal Conditioning Amplifi er System frame.
4. Locate the 3-pin jumper J16. Jumper pins 1 and 2 to use the SI-BAM21-LCB with X1
gain multiplier, or jumper pins 2 and 3 for use with X10 gain multiplier.
5. Reinstall the module into the frame and secure it with the screws.
Using the Anti-Oscillation Unit
Adjusting the Anti-Oscillation Filter
The anti-oscillation fi lter is adjusted at the factory using the transducer that is supplied
with the SI-MT, SI-MKB, SI-HTB, or SI-CTS system. Normally, the fi lter does not need to
be reset, unless a different force transducer is connected to the unit. To adjust the anti-
oscillation fi lter properly, the transducer is excited at its resonance frequency using a
magnetic driver or pulser (WPI #97204).