Instruction manual
SI-H Muscle Testers
16
WORLD PRECISION IN STRU MENTS
Confi guration Buttons–The Display button is used to toggle the display between Channel
1 temperature, Channel 2 temperature and both. The Setup button rotates through the
array of confi gurable parameters. The Up and Down buttons are used to adjust the
parameters.
Cuvette Connections–Use these ports to connect SI-H cuvettes used with the SI-MT and
SI-MKB platforms.
SI-COLUB
The SI-COLUB Constant Load Module can be used in one of three different modes. In
its primary mode (Constant Load), the unit can perform constant load experiments. In its
primary mode (Constant Load) the unit takes an external trigger command from the force
transducer to perform a constant load cycle. In addition, the module allows for a different
external trigger, or you can completely bypass the module without having to switch
cabling.
The SI-COLUB Constant Load Module can be used to maintain a constant force on a
muscle, muscle length or sarcomere length rather than the constant length of the muscle
preparation. This is accomplished using a feedback loop. The SI-COLUB is set to monitor
a specifi c parameter, like force or sarcomere length, and it uses that value to create a
motor position control signal that adjusts the position of the motor. The SI-COLUB also
receives a feedback signal that it uses to create control signals that constantly adjust the
position of the motor until the commanded setpoint is reached.
Through MDAC software, the LabTrax 8/16 can be programmed to create position,
stimulation, and command protocols for muscle experiments. In simple experiments,
the protocols would move the motor to stretch or release a muscle and stimulate it to
contract. In more complex experiments, MDAC and the LabTrax 8/16 might be used to
control a module like the SI-COLUB Constant Load Module that monitors the load on the
muscle or the length of its sarcomeres.
Pulse Train from LabTrax 8/16
F0
Force
Force Output of the Tissue Sample
Slack Period Restretch
Isometric Force
Develo
p
ment
kF0
Voltage
Fig. 17 The digital output from the LabTrax 8/16 data acquisition unit sends a capture
command to the SI-COLUB to slack or restretch the muscle as it is maintained under load.