Data Acquisition Output–Connect a data acquisition system like WPI’s Lab-Trax-8/16 to
this BNC connector to record the raw SI-BAM21-LCB voltage output. For test purposes, a
multi-meter or oscilloscope may be connected using a standard BNC cable (WPI #2851).
Force Transducer Connection–An SI-KG force transducer is plugged into this DIN
connector. Align the pins, and insert the connector until it is fully seated.
Digital Interface–This connection is a legacy interface for classic SI-H equipment.
NOTE: When the SI-MT or SI-MKB electronics are confi gured at the factory for the
muscle Tester systems, the signal is routed internally from the SI-BAM21-LCB module to
the SI-AOSUB module. The Force Output connection on the front of the SI-BAM21-LCB
module also shows the raw unfi ltered signal from the transducer, but it does NOT need to
be connected externally.
SI-AOSUB
Every force transducer has a resonance frequency at which it vibrates. The SI-AOSUB
allows you to locate that frequency and fi lter the signal to mitigate the noise of the
resonance frequency. Since each force transducer is unique, the anti-oscillation unit must
be calibrated for each force transducer. Likewise, the tissue mounting hardware affects
the resonance frequency. Therefore, the system must be calibrated with the mounting
hardware attached to the force transducer.
Front Panel
Pulser Amplitude Adjustment Knob
Pulser Port
External Input Port
Anti-Oscillation Frequency
Adjustment Potentiometer
Signal Amplitude
LED Array
Filtered Signal Output Port