1) Removal of the Top Cover
Remove the top cover screws (2 along upper front panel, 2 per each side, 3 along upper rear panel). The 919126 PCB is vertically
mounted to chassis bottom just behind the front panel mounted LED driver PCB (see photo below).
2) Adjustment Pot Locations and Functions
On the 919126 PCB(s) the P2 pot adjusts the left (Channel A) bargraph meter and the P3 pot adjusts the right (Channel B) bargraph
meter. See diagrams below.
3) “0” Fine Adjustment Setup
Connect a test signal of approximately 1 kHz of the amplitude desired to give a “0” indication on the bargraph. This signal should
be connected across the balanced inputs, and should be from a balanced source.
4) Fine Adjustment of the Bargraph “0” Level
Turning each pot clockwise increases the displayed level (sensitivity) of the associated bargraph meter. For example, to set
the zero level at +7 dBu (rather than the factory default setting of +4 dBu), adjust the appropriate DIP switch for gain closest
to that desired (see page 22 for Gain Calibration settings). Feed a +7 dBu signal (see step 3) into the analog input on the rear
panel (Item F, page 16), then turn the pots counter-clockwise as you visually monitor the associated bargraph meter on the front
panel until the desired display setting (“0”) is achieved for the increased input level. To zero the meters for a lower input level,
you would turn the pots clockwise to increase the displayed level for any decreased input level.
Note: For range setting, use the Meter Input Gain Calibration/Bargraph Display Mode DIP Switches
Level Meter Bargraph "0" (zero) Fine Adjustment
P2 pot adjusts the left bargraph meter.
P3 pot adjusts the right bargraph meter.
Turn pots clockwise to
increase the display level (sensitiv-
ity) in the associated bargraph meter
Section 3: Technical Information
R10
R12
IC5
P2
P3