User guide
821024: AMP1-DA/106 User Guide
© 2011 Wohler Technologies, Inc. All rights reserved.
16
AMP1-DA/106 User Guide
Rear Panel 10-Position DIP Settings for the Level Meters
7. Only one auto-calibration attempt may be made for each cycling of AC
power to the unit. Once the Line Level Calibration DIP switch has been
placed in the
CAL position, it is necessary to cycle the power before that
DIP switch will be functional again, even if a calibration attempt
was unsuccessful. If one wishes to calibrate again, turn off the
power to the unit and repeat steps 1 through 6.
Reference Level
DIP switch sections 2 and 3 determine the Reference Level, which
adjusts the level of the input signal and the resultant level displayed on
the LED bargraphs. Factory setting is +4 dB. See Figure 1–8 on page 14
for settings.
Bargraph Display Mode
DIP switch sections 4 and 5 determine how peak levels are displayed
for the associated meters on the front panel. There are four possible
settings:
• VU Only,
• VU-PPM Floating Segment,
• PPM Only, and
• PPM-PPM Floating Segment.
The VU Only selection has a VU floating segment when a Peak Hold
value is selected using the Internal 10-Position DIP Switch Module. The
factory default setting is VU-PPM Floating Segment. See Figure 1–8 on
page 14 for settings.
Rear Panel 10-Position DIP Settings for
the Level Meters
This 10-position DIP switch is accessed by removing the top cover of
the AMP unit and is located on the PCB (the same PCB on which the 6-