Rear Panel 6-Position DIP Settings for the Level Meters
Figure 1–8 AMP1-DA/106 Rear Panel DIP Switch
Settigns
Line Level (Auto) Calibration
The unit is calibrated at the factory. To recalibrate:
1. Turn on the power.
2. Apply the desired reference level (nominal 0) signal to all channels.
3. Make sure the Reference Level DIP sections (2 and 3) are set to the
nearest level of the input signal being applied for calibration (i.e., 0, +4,
+6 or +8). The user should make sure that the signal applied to all
four channels is within +/- 4 dB of the reference level selected by DIP
switch sections 2 and 3.
4. Place DIP section 1 in the down position.
5. Wait 10 seconds. The unit will remove the previous calibration and the
new calibration will be applied.
6. Place DIP section 1 in the up position and return unit to service.
Note:
For more accurate indication of signal levels, meters are
tuned to effect a “rounding” function, which occurs between
the thresholds of any two bargraph segments. This means
the level meter zero LED segment will turn on at one-half
the smallest spacing between LED segments (mid-scale
resolution) before that segment's scale indication.