Regulatory information
Guidance and manufacturer’s declaration-electromagnetic
emissions
This cellular Smart Blood Pressure Monitor is intended for use in
the electromagnetic environment specified below. The customer
or the user of the cellular Smart Blood Pressure Monitor should
ensure that it is used in such an environment.
Emissions test Compliance
Electromagnetic environ-
ment — guidance
CE emissions
CISPR11
Group 1
The cellular Smart Blood
Pressure Monitor uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RE emissions
CISPR11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
This cellular Smart
Blood Pressure Monitor
is suitable for use in all
establishments, including
domestic establishments
and those directly
connected to the public
low-voltage power supply
network that supplies
buildings used for
domestic purposes.
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Not applicable
Surge IEC
61000-4-5
1 kV differential
mode
2 kV common
mode
ine(s) to
line(s):
±1 kV
100 kHz
repetition
frequen-
cy
The main power quality
should be similar to that
of a typical commercial
or hospital environment.
The cellular Smart Blood Pressure Monitor declaration
electromagnetic immunity
This cellular Smart Blood Pressure Monitor is intended for use in
the electromagnetic environment specified below. The customer
or the user of the cellular Smart Blood Pressure Monitor should
ensure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic envi-
ronment – guidance
Conducted
RFIEC 61000-
4-6
3 Vrms
150 kHz to
80 MHz
N/A N/A
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
N/A Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
equipment or system,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter. Interference
may occur in the vicinity
of equipment marked
with the following
symbol.
Electrostatic
discharge
(ESD)IEC
61000-4-2
Contact: ±8 kV
Air: ±2 kV, ±4
kV, ±8 kV,
±15 kV
Contact:
±8 kV
Air: ±2
kV, ±4 kV,
±8 kV,
±15 kV
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/
output lines
N/A The main power quality
should be similar to that
of a typical commercial
or hospital environment.