Error messages
16
Error message Causes Countermeasures
Check that the Blood Pressure Monitor is
correctly positioned on your arm and that mea-
surement is performed in good conditions.
- User movement, or not relaxed, or talking …
- Cuff is not correctly fastened
- Cuff broken
- Pump or valve failure
- Pressure overflow
- Noise interference
- User movement, or not relaxed, or talking …
- User special characteristics
- Battery level is too low
1. Check that the Blood Pressure Monitor
is correctly positioned on your arm and that
measurement is performed in good conditions.
2. Low battery level. Load your BPM Connect
with a micro USB cable.
3. If user has special characteristics, please
contact your physician.
4. If the problem occurs again, please contact
customer service.
Low battery level. Battery level is too low. Load the device with the provided charging
cable.