Owner's Manual
Table Of Contents
- Quick Installation Guide
- Information Guide
- Indications for use
- Warnings
- Important Notes
- Using the Scan Monitor
- What is an ECG?
- Taking an ECG
- ECG Outputs
- Sharing ECG Results
- What is an SpO2 Measurement?
- Taking an SpO2 Measurement
- SpO2 Outputs
- SpO2 Measurement Information
- Technical specifications
- Wireless information
- Security
- Safety and Performance
- Troubleshooting - Syncing with App
- Troubleshooting - ECG
- Troubleshooting - SpO2
- Cleaning and Disposal
- Equipment symbols
- Warranty
Safety
and
Performance
36
Declaration:
Electromagnetic
emissions
and
immunity
for
equipment
and
systems
that
are
not
life-supporting
and are
specified
for
use
only
in a
shielded
location.
Declaration:
Electromagnetic
immunity
Scan
Monitor
is
intended
for
use in
the
electromagnetic
environment
specified
below. The
customer
or
the
user
of
the
Scan
Monitor
should ensure
that
it
is used in such an
environment.
Immunity test
Conducted
RFIEC
61000-4-6
Radiated RF
IEC
61000-4-3
Electrostatic discharge
(ESD) IEC
61000-4-2
IEC
60601
Test Level
3 Vrms
150kHz
to
80MHz
l0V/m
80
MHz
to
2.7 GHz
Contact:
±2kV,
±4kV, ±6kV, ±
8kV
Air:
±2kV, ±4kV,
± 8kV, ±15kV
Compliance level
N/A
N/A
N/A
Contact: ±2kV,
±4kV, ±6kV, ±
8kV
Air: ±2kV, ±4kV,
± 8kV, ±15kV
Electromagnetic environment-guidance
N/A
Portable and
mobile
RF
communications
equipment
should
be
used
no
closer
to
any
part
of
the
equipment
or
system including cables than
the
recommended separation distance calculated
from
the equation applicable
to
the frequency
of
the
transmitter. Interference may occur in the
vicinity
of
equipment
marked
with
the
following symbol
Floors should be wood, concrete
or
ceramic tile.
If
floors are covered
with
synthetic material, the
relative
humidity
should be
at
least 30%.