COATING THICKNESS GAUGE
INSTRUCTION MANUAL
3
1 42
A.Introduction
⑤
⑥
①
(Figure 1)
OK
OK
③
④
②
⑦
⑧
Measurement interface
(Figure 2)
a
c
e
f
g
i
k
b
a
h
j
l
d
f
This product is a color-screen portable coating thickness gauge with high-
definition display, which can quickly, non-destructively and accurately me-asure
non-magnetic coating thickness on magnetic metal substrates and non-metallic
coating thickness measurement on non-magnetic metal sub-strates. At the same
time it can automatically identify magnetic metal substrate and non-magnetic
metal substrate, and is widely used in manufacturing, metal processing industry,
chemical industry, commodity inspection and other testing areas.
B.Functions
Menu operation and color-screen HD display.
Thickness measurement of non-magnetic coating on magnetic metal
substrate surface and non-metallic coating on non-magnetic metal
substrate .
Two measurement methods: single measurement, continuous
measurement.
Basic calibration and Zero-point calibration available.
Metric/imperial unit and storage function.
Screen rotation, charge protection, multi-interface displays, screen
brightness selection.
Automatic shutdown.
C.Name of Parts(Such as Figure 1)
①.USB charging interface
②.LCD display
③.Back button and power on/off button
④.Confirm button
⑤.Up button
⑥.Down button
⑦.Test probe
⑧.Nameplate on instrument’s back
D.Operation manual
1.Measurement instruction: After startup, lightly press test probe on the
substrate to be tested under measurement interface, and the measured
value is coating thickness of substrate.
2.Measurement interface (Such
as Figure 2):
a. Mode: single (sng) /
continuous (ctn)
b. Battery power indicator
c. Measurement counting
d. Substrate: Fe/nFe
e. Display zone of measured
value
f. Unit : um/mil
g. Difference value of
measurement
h.Average value
. Maximum value
. Minimum value
k. High limit l. Low limit
3.Enter menu interface:
▶Under measurement interface, short press “ ” button to enter menu
interface, which displays button prompt afterwards (Figure 3):
▶Press “OK” button under measurement interface for quick entry of measured
data menu (Figure 4):
4. Screen rotation (Such as Figure 5 ~ Figure 6):
▶ Long press button under
measurement interface to rotate
screen.
I
j
“ ”
Enter menu interface
(Figure 3)
Enter menu interface
(Figure 4)
Horizontal screen display
(Figure 5)
Vertical screen display
(Figure 6)
5.Functions in Menu (Such as Figure 7):
▶Measurement selection: Enter measurement sub-items to operate.
Functions in Menu (Figure 7)
1. Calibration Operation
(1).Basic calibration:
a. Prepare the calibration plate and calibration base, enter calibration menu
and select basic calibration.
b. According to the instrument instruction, place the corresponding calibration
plate for calibration.
c. After calibration is completed, “calibration complete” will show up at the
bottom of the screen and the instrument will return to the previous interface.
d. After the calibration is complete, you can go back to measurement interface
and perform measurement.
e.Magnetic and non-magnetic metal base calibration do not affect each other
(2).Zero Calibration:
a. Enter calibration menu and select zero calibration.
b. Lightly press the instrument onto substrate.
c. The instrument will automatically calibrate to zero point.
d. The instrument displays the previous interface after calibration is done.
e. After calibration is complete, you can go back to measurement interface
and perform measurements.
7. Limit value:
▶If limit value switch is on, when the measured value is above
high limit or below low limit, the value will turn red and the
corresponding icon will appear (Such as Figure 8).
▶If limit value switch is off, value of normal measurement will
be displayed (Such as Figure 9).
8. Storage (Such as Figure 10):
▶Measured data: users can view, save and delete measured data.
▶View: view saved data (Such as Figure 11).
▶Delete: under delete interface, delete saved data by short pressing OK
button to delete single datum, or long pressing OK button to select all
data and delete.
Note: When performing basic calibration, press the instrument onto substrate
when the arrow is down, and remove it when the arrow is up.
1.Language settings >>
2.Automatic shutdown >>
3.Voice prompt (on/off)
4.High backlight (on/off)
5.Selected colors >>
6.Software version >>
7.Restore factory settings >>
Setting
1.Chinese
2.English
Language
1.Two minutes
2.Five minutes
3.Ten minutes
4.Halfhour
5.Never
No operation
auto off time
1.Measure mode >>
2.Measure unit >>
Measure
1.Single(SNG)
2.Continuous (CTN)
Measure mode
1.mil
2.um
Measure unit
1.black
2.Blue
3.Orange
Selected Colors
View
1.um & mil
2. Measurement statistics
3. Past values of measurement
4. Measurement trend chart
(on/off)
(on/off)
(on/off)
(on/off)
1.Measure >>
2.Calibration >>
3.Limit >>
4.Memory >>
5.View >>
6.Settings >>
Menu
1.Base calibration >>
2.Zero calibration >>
Calibration
1.Limit value (on/off)
2.High limit >>
3.Low limit >>
Limit
1. Measured data >>
2. view group >>
3. Delete group >>
Memory
1. View data >>
2. Save data >>
3. Delete data >>
Measured data
7
5
6
8
DC 3.7V (Lithium battery capacity 1000mAh)
Menu Chart (Figure17)
Above high limit
(Figure 8)
Normal measurement display
(Figure 9)
Storage (Figure 10) Measured data (Figure 11)
9. View:
▶Turn on um&mil switch and measured results of two units are
simultaneously displayed under measurement interface (Such
as Figure 12).
▶Turn on other switches and press UP/DOWN button under
measurement interface to view the corresponding interface (Such
as Figure 13~15).
Interface with Um & mil on (Figure 12)
Measurement statistics
(Figure 13)
Past values of measurement
(Figure 14)
Measurement trend chart
(Figure 15)
Settings
(Figure 16)
2mm
Measurement range
Resolution
Minimum diameter of magnetic metal substrate
Minimum thickness of magnetic metal substrate
Minimum radius of curvature for magnetic convex substrate
( 0.1μm( <100μm) ,1μm( ≥100μm) ) /0.1mil
0.5mm
12mm
>150μm ±( 3%H+1μm)
≤150 ± μm 5μm
0 ~ 1300um/51mil
Measurement error
11mm
Minimum radius of curvature for magnetic concave substrate
0.5mm
Minimum thickness of non-magnetic metal substrate
50mm
Minimum diameter of non-magnetic metal substrate
E.Technical Parameters
Working voltage
Reminder: This instrument is equipped with rechargeable battery. If you can't
turn on the product, please use it after charging (power adapter
specification: DC 5V/1A, namely ordinary mobile phone charger).
This instrument is not shipped with power adapter.
F.Menu Chart(Such as Figure17)
1.Keep the probe away from the
measured substrate when starting up.
2.“- OL-”indicates that the value exceeds
measurement range.
3.Do not press the sensor with your finger or
other object into the instrument, because
this operation may damage the sensor
parts and the instrument.( Figure 18)
G.Attention
Specific Declarations:
Our company shall hold no any responisibility resulting from using
output from this product as an direct or indirect evidence.
We reserves the right to modify product design and specification
without notice.
52.9*26*117mm
Size
Weight
102.4g( including battery)
10. Settings:
▶Enter setting menu to perform corresponding operations.(Such as Figure 16).