OWNER’S OPERATING & MAINTENANCE MANUAL S300 Stretchair S550, S750, & S999 Bariatric Stretchair PLEASE DO NOT DISCARD THIS MANUAL! KEEP FOR FUTURE REFERENCE AND TRAINING
TABLE OF CONTENTS/WARRANTY IMPORTANT SAFETY INFORMATION.........................................................................................2-7 CHAIR SPECIFICATIONS.............................................................................................................8-9 OPERATING INSTRUCTIONS.............................................................................................….....10-14 GENERAL MAINTENANCE...................................................……................................
IMPORTANT - PLEASE READ Read all instructions before using this product. IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS To reduce the risk of electric shock when using an electrical appliance, basic precautions should always be followed, including the following: DANGER: Always unplug this product from the electrical outlet before cleaning, maintenance or when putting on or taking off parts. WARNING: 1.
IMPORTANT - PLEASE READ 1. READ AND FOLLOW ALL DIRECTIONS. 2. These chairs are not to be used in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide. 3. DO NOT allow children or pets to play on, under, or around the Stretchair. 4. DO NOT put hands, feet or clothing into any openings when changing positions or adjusting arms/legrests on chair. Attendant MUST confirm that users arms, legs, hands and feet are safe while changing chair positions or SERIOUS INJURY MAY RESULT. 5.
IMPORTANT - PLEASE READ Products with the following markings are intended for commercial use only. Independent U.S.
IMPORTANT - PLEASE READ 1 Guidance and manufacturer’s declaration –electromagnetic emissions 2 The Stretchair is intended for use in the electromagnetic environment specified below. The customer or the user of the Stretchair should assure that it is used in such an environment.
IMPORTANT - PLEASE READ Guidance and manufacturer’s declaration –electromagnetic immunity The Stretchair is intended for use in the electromagnetic environment specified below. The customer or the user of the Stretchair should assure that it is used in such an environment.
IMPORTANT - PLEASE READ Guidance and manufacturer’s declaration –electromagnetic immunity The Stretchair is intended for use in the electromagnetic environment specified below. The customer or the user of the Stretchair should assure that it is used in such an environment.
CHAIR SPECIFICATIONS S300 S550 28"...........................Overall Width..............................34" 53.5"........................Overall Height.............................53.5" 22"...............................Seat Height.............................22" 19.5.............................Seat Width...............................25" 72"............................Layflat Length............................72" 22"-38"........................Vertical Lift (Chair)......................22"-38" 22"-38".
OPERATING INSTRUCTIONS Assisting Patient Into Stretchair 1. Stretchair should be in upright position. (FIG. 1) 2. Apply central locking wheel brakes (S675 & S999) or all 3 total lock casters (S300). (FIG. 2a-2b) 3. Adjust armrests to the DOWN position. (SEE PAGE 13) 4. Pivot the footrests to upright position with the padded surface facing outward. (FIG. 1) 5. Assist patient into chair and secure them with the safety belt provided. 6.
OPERATING INSTRUCTIONS Converting Layflat Position to Chair Lock wheels (FIG. 2A or 2B) Keep armrest/side-rails in the fully upright position. (SEE PAGE 13 FIG.3) Secure patients arms in front of their chest Using the handset control, return the chair to its sitting position. (FIG. 1) Raise the patient's legs and lower the footrests. Adjust to a suitable height for the patient. (SEE PAGE 12) 6. Armrests can now be lowered to a comfortable height for the patient. 7.
OPERATING INSTRUCTIONS SWING-LEG OPERATING INSTRUCTIONS The swing-leg is designed to elevate one or both patient's legs to a comfortable or prescribed position. Once the leg is elevated it will remain at a fixed angle, relative to a patient's thigh, regardless of the chair angle. Desired positioning of legs can be achieved with 4 length settings and infinite elevation adjustments. In the layflat position, the swing legs are NOT adjustable. Adjusting Swing-leg for Patient Comfort 1.
OPERATING INSTRUCTIONS ADJUSTABLE ARMREST INSTRUCTIONS The Stretchair armrests are adjustable to enhance the patients comfort while sitting (FIG. 7), and act as guard-rails while the chair is in the layflat position. (FIG. 9) Caution: DO NOT allow patients to use the stand assist grips or attempt to exit the chair until the armrests have been adjusted to their lowest position, and the wheels have been locked. Adjusting Armrests for Patient Comfort 1.
OPERATING INSTRUCTIONS LINAK BATTERY & CONTROL BOX INSTRUCTIONS • To charge battery BE SURE the emergency stop is in the "OUT" position. *Battery WILL NOT charge if the emergency stop is engaged. • Keep patients away from Stretchair while it's charging. • Battery MUST BE fully charged prior to FIRST use. • Without a battery back-up DO NOT allow chair to run with less than a 50% charge. • Battery cannot be over-charged using the charging unit provided and the instructions given. FIG. 10 FIG.
MAINTENANCE GUIDE Monarch Hydro Power Unit Replacement 1. Place chair in the layflat position (SEE PAGE.7) with the vertical lift in its LOWEST position. (FIG.15) 2. Unplug the Monarch Power cord from the battery control box. (FIG.16) 3. Locate the Power Unit positioned under the upholstered seat. (FIG. 17) 4. Remove the 5 screws and ALL nylon ties that ` secure the Power Unit to the chair frame. 5. Lift, rotate, and remove unit from the chair frame.
MAINTENANCE GUIDE Linak Actuator Replacement 1. Place chair in the layflat position (SEE PAGE 7) with the vertical lift in its highest position. (FIG.20) 2. Unplug the Linak Actuator cord from the battery control box. (FIG.21) 3. Locate the Linak Actuator positioned just behind the Monarch Power Unit. (FIG.23) 4. Cut all nylon ties that secure the Linak Actuator's power cord to the chair frame. 5. Remove the upper mounting bolt and spacers. (FIG.22) 6.
MAINTENANCE GUIDE Swing-leg Gas Spring Replacement 1. Fully elevate the legrests for easier access to the gas springs. (FIG.24) 2. Remove the swing-leg release handle by unscrewing the hex-head bolt. (FIG.24) 3. Locate the 2 gas springs positioned above the Monarch Power Unit 4. Remove (A) the front mounting pin and spacers first and then (B) the rear mounting pin and spacers. Keep for re-use. (FIG.25) 5. Discard old gas spring 6.
OUR PRODUCTS Other Quality Winco Furnishings You May Be Interested In: Visit us online at www.wincomfg.
LIMITED WARRANTY TERMS Winco Mfg., LLC. warrants this product to be free of manufacturer’s defects in material and workmanship, provided that the product is used according to normal operating conditions and proper maintenance intended by manufacturer. This warranty is available only to the original retail purchaser, is non-transferable and commences on the date of retail sale; proof of purchase required. Limited One Year Complete Product Warranty: Winco Mfg., LLC.