User's Manual
Chapter 3: Programming and Calibration
Exit Sub-Block
(12/00) 3-9
[F3.3 2] STOP BITS
1 = 1 stop bit
2 = 2 stop bits
[F3.4 2] PARITY
0 = No parity
1 = Odd parity
2 = Even parity
[F3.5 2] DATA OUTPUT FORMAT
0 = Mettler Toledo Continuous with STX
1 = Demand, single line, displayed weight only
2 = Demand, single line, gross, tare, net
3 = Demand, three line gross, tare, net
[F3.6 1] CHECKSUM (Only if F3.5 = 0)
0 = No checksum
1 = Checksum
[F3.7 0] PRINTED GROSS WEIGHT LEGEND
0 = No Legend
1 = B (bruto)
2 = G (gross)
Exit Sub-Block
At the end of the programming mode, there are three choices to exit. Use the FUNCTION
key to move between the choices.
[SAVE ] Press PRINT to accept the changes in the program mode and exit
programming.
[Abort ] Press PRINT to ignore any changes in the program mode and exit
programming.
[dFAULT] Press PRINT to reset all program block parameters to factory defaults and
exit programming.
Parameters marked with an asterisk (*) in Table 3.1 will not be reset