User's Manual
METTLER TOLEDO WILDCAT Scale Technical Manual
(12/00) 3-8
[F2.6 1] MOTION SENSITIVITY SELECTION
The motion detection feature determines when a no-motion condition
exists on the scale platform. The sensitivity level determines what is
considered stable. Printing, pushbutton zero, and tare entry are blocked
during instability.
0 = Motion detector disabled
1 = 1.0 d motion sensitivity
2 = 3.0 d motion sensitivity
[F2.7 0] FILTER
Filtering compensates for environmental disturbances such as vibration
or drafts.
0 = NONE
1 = LIGHT
2 = NORMAL
3 = HEAVY
[F2.8 0] SLEEP MODE (power saving)
0 = Disable
1 = Enable the sleep mode automatically after 5 minutes of stability.
[F2.9 1] POWER UP ZERO RANGE
0 = Auto Zero capture at powerup disabled.
1 = Auto Zero capture at powerup range of +/-2%
2 = Auto Zero capture at powerup range of +/-10%.
If power-up zero capture is disabled, the scale will display weight after
powerup. Otherwise, if weight is not in zero-capture range, the display
shows [E E E] or [-E-E-E], until weight is within the capture range.
F3 Interface Block
[F3.1 ] BAUD RATE
[ 9600] Choose a baud rate from a selection list of 1200, 2400, 4800, or 9600
baud
[F3.2 7] DATA BITS
7 = 7 data bits
8 = 8 data bits