User's Manual
Chapter 3: Programming and Calibration
F2 Application Block
(12/00) 3-7
F2 Application Block
[F2.1 0] ALTERNATE UNITS:
Select the unit of measure desired as a secondary unit.
0 = No unit switching
1 = lb
2 = kg
If the calibration unit is “kg ”, the available choice is only “lb”. If the
calibration unit is “lb” or ‘g’, the choice is only ‘kg”. If unit switching is
enabled, a quick press of the FUNCTION key will change the unit.
[F2.2 0 ] AUTO BACKLIGHT
0 = Backlight can only be turned on manually by pressing the
FUNCTION key.
1 = The backlight turns on during motion and stays on for 6 seconds
after no-motion.
The manual on/off is always available. If unit switching is enabled, press
and hold the FUNCTION key for 3 seconds to turn the back light on. If
unit switching is disabled, a quick press of the FUNCTION key will turn
the backlight on and off.
[F2.3 1] TARE
0 = Tare disabled
1 = Tare enabled
[F2.4 1] PUSHBUTTON ZERO RANGE
0 = Pushbutton zero disabled
1 = Enable pushbutton zero within +/- 2% of scale capacity
2 = Enable pushbutton zero within +/- 20% of scale capacity
[F2.5 1] AUTO ZERO MAINTENANCE
Auto Zero Maintenance (AZM) automatically compensates for small
changes in zero resulting from material build-up or temperature changes.
Select the weight range (+/-) around gross zero within which the scale
will capture zero. If residual weight on the scale exceeds the weight
range, the scale will not capture zero. AZM is disabled in the net mode.
0 = No AZM.
1 = AZM within 0.5d window.
2 = AZM within 1d window.
3 = AZM within 3d window.