WILDCAT Scale Technical Manual E15609900A 12/00.
© Mettler-Toledo, Inc. 2000 No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording, for any purpose without the express written permission of Mettler-Toledo, Inc. U.S. Government Restricted Rights: This documentation is furnished with Restricted Rights.
CUSTOMER FEEDBACK Your feedback is important to us! If you have a problem with this product, or just a suggestion on how we can serve you better, please fill out this form and send it to us. If you are in the United States, you can mail this postpaid form to the address on the reverse, or fax it to (614) 438-4355. If you are outside the United States, please apply the appropriate amount of postage before mailing. You can also send your feedback via email to: quality_feedback.mtwt@mt.com.
FOLD THIS FLAP FIRST NO POSTAGE NECESSARY IF MAILED IN THE UNITED STATES BUSINESS REPLY MAIL FIRST CLASS PERMIT NO. 414 COLUMBUS, OH POSTAGE WILL BE PAID BY ADDRESSEE Mettler-Toledo, Inc. Quality Manager - MTWI 1150 Dearborn Drive Worthington, Ohio 43085 USA Please seal with tape.
INTRODUCTION Information regarding METTLER TOLEDO Technical Training may be obtained by contacting: METTLER TOLEDO 1900 Polaris Parkway Columbus, Ohio 43240 (US and Canada) 614- 438-4511 (All Others) 614-438-4888 FCC Notice This device complies with Part 15 of the FCC Rules and the Radio Interference Requirements of the Canadian Department of Communications.
PRECAUTIONS SAFETY NOTICE Product safety is a fundamental concern at METTLER TOLEDO. Use common sense and follow the simple precautions listed below to ensure your safety and optimize the use and performance of this product. • • • • • • • • • • Read this manual before operating or servicing this product. Save this manual for future reference. Observe safety warnings located throughout this manual. Use caution when lifting or moving heavy equipment.
CONTENTS 1 Introduction................................................................................................... 1-1 Specifications..................................................................................................................... 1-1 2 Assembly Procedures..................................................................................... 2-1 Assembly Instructions .........................................................................................................
5 Parts List ...................................................................................................... 5-1 Column and Base (WS30, WS60, WS150)........................................................................... 5-2 Column and Base (WW250 and WW500) ............................................................................ 5-3 6 Appendices ................................................................................................... 6-1 Appendix 1: Demand Mode Output............
Chapter 1: Introduction Specifications 1 Introduction The WILDCAT bench and portable scale is designed to meet the real world requirements of manufacturing, agricultural, packaging, and general weighing applications, and is ready to use right out of the box. This manual provides important information for installing and servicing the scale. For information on operating the scale, refer to the Operator’s Card provided with the unit.
METTLER TOLEDO WILDCAT Scale Technical Manual WS30R/WS30VR WS60R/WS60VR WS60LR/WS60LVR Displayed Resolution WS150R/WS150VR WW250VR WW500VR Up to 10,000d Maximum Capacity Minimum Division Size Trade -Approved Build Maximum Back weight Safe Overload 30 kg (60 lb) 60 kg (150 lb) 60 kg (150 lb) 150 kg (300 lb) 250 kg (500 lb) 500 kg (1000 lb) 0.01 kg (0.02 lb) 0.01kg (0.05 lb) 0.02 kg (0.05 lb) 0.05 kg (0.1 lb) 0.1 kg (0.2 lb) 0.2 kg (0.
Chapter 2: Assembly Procedures Assembly Instructions 2 Assembly Procedures Assembly Instructions Assembly instructions for the WILDCAT scale are provided with the unit. A copy of these instructions appears below and on the next page.
METTLER TOLEDO WILDCAT Scale Technical Manual Scale Assembly Instructions Báscula Portátil de Mesa Instrucciones Para Ensamblaje Instructions D‘assemblage de la Bascule Additional tools you may need Herramientas adicionales que pudiese necesitar Outils supplémentaires dont vous pourriez avoir besoin Plug in or use 6 “ D ” size batteries Enchufa o utilice seis baterías tipo “D” Branchez le courant ou utilisez 6 piles de taille “D” © 2000 Mettler-Toledo, Inc.
Chapter 2: Assembly Procedures Load Cell Wiring Load Cell Wiring Figure 2-3 shows the load cell terminal strip wiring for the WILDCAT scale. +EXC Green +SEN Blue +SIG Red Shield Orange -SIG White -SEN Brown -EXC Black Figure 2-3 Serial Port Connections The WILDCAT scale provides a bi-directional RS-232 port to be used to send data to or receive commands from a computer, printer, or other serial device.
METTLER TOLEDO WILDCAT Scale Technical Manual NOTES 2-4 (12/00)
Chapter 3: Programming and Calibration Serial Port Connections 3 Programming and Calibration WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT IN BODILY HARM. The WILDCAT scale is calibrated at the factory. Functional programming is also completed at the factory, so that the WILDCAT scale is ready to use right out of the box.
METTLER TOLEDO WILDCAT Scale Technical Manual Key Functions Should you need to reconfigure the WILDCAT scale, the following keys are used in the programming mode to configure the program blocks. 0 T C F ZERO Back up to the previous step. TARE Moves the data entry position one digit to the left. FUNCTION Increments the numeric data entry digit and/or allows the programmer to view the next display in a selection list. PRINT Accepts/terminates a data entry.
Chapter 3: Programming and Calibration Entering and Configuring Program Blocks Entering and Configuring Program Blocks Once the [F1] prompt is displayed, use the PRINT key to enter the block. Use the FUNCTION key to skip to the next block. Use the ZERO key to go to the previous block. Once PRINT is pressed, the configuration advances to the first parameter in the F1 program block. The display shows the sub-block number and the current value setting.
METTLER TOLEDO WILDCAT Scale Technical Manual Factory Default Settings STEP DEFAULT DESCRIPTION F1.1 *, ** Calibration units – No default GEO * Gravity adjust – No default F1.2 0 Skip calibration F1.3 0 Normal weight display F1.4 0** Programming mode access disabled F2.1 0** Alternative units = none (Unit switch disabled) F2.2 0** Auto Backlight disable F2.3 1 Tare enable F2.4 1** Push button zero enabled, 2% range F2.5 1** Auto zero maintenance enabled within 0.
Chapter 3: Programming and Calibration F1 Scale Block F1 Scale Block The scale program block allows the user to set and calibrate the features that affect weighing performance. [F1.1 2 ] CALIBRATION UNITS Enter the value that corresponds to the type of test weights that will be used for calibration. 1 = lb 2 = kg 3=g [ GEO 12 ] GRAVITY ADJUST The WILDCAT scale is calibrated with a GEO code of 12 at the factory.
METTLER TOLEDO WILDCAT Scale Technical Manual [Incr. ] INCREMENT SIZE “Incr” displays momentarily then the current increment size is displayed for editing. Press the FUNCTION key to toggle through valid selections. [E SCAL] Empty scale platform and press PRINT to continue. [15 CAL] Display countdown for zero setting. If motion sensitivity is not disabled and motion is detected at this step, the display will show [ E 30]. Press PRINT and the display returns to the [E SCAL] prompt.
Chapter 3: Programming and Calibration F2 Application Block F2 Application Block [F2.1 0] ALTERNATE UNITS: Select the unit of measure desired as a secondary unit. 0 = No unit switching 1 = lb 2 = kg If the calibration unit is “kg ”, the available choice is only “lb”. If the calibration unit is “lb” or ‘g’, the choice is only ‘kg”. If unit switching is enabled, a quick press of the FUNCTION key will change the unit. [F2.
METTLER TOLEDO WILDCAT Scale Technical Manual [F2.6 1] MOTION SENSITIVITY SELECTION The motion detection feature determines when a no-motion condition exists on the scale platform. The sensitivity level determines what is considered stable. Printing, pushbutton zero, and tare entry are blocked during instability. 0 = Motion detector disabled 1 = 1.0 d motion sensitivity 2 = 3.0 d motion sensitivity [F2.7 0] FILTER Filtering compensates for environmental disturbances such as vibration or drafts.
Chapter 3: Programming and Calibration Exit Sub-Block [F3.3 2] STOP BITS 1 = 1 stop bit 2 = 2 stop bits [F3.4 2] PARITY 0 = No parity 1 = Odd parity 2 = Even parity [F3.5 2] DATA OUTPUT FORMAT 0 = Mettler Toledo Continuous with STX 1 = Demand, single line, displayed weight only 2 = Demand, single line, gross, tare, net 3 = Demand, three line gross, tare, net [F3.6 1] CHECKSUM (Only if F3.5 = 0) 0 = No checksum 1 = Checksum [F3.
METTLER TOLEDO WILDCAT Scale Technical Manual Calibration When the WILDCAT scale is used in legal-for-trade commercial applications, it must be calibrated with certified test weights to the capacity and increment size shown on the data plate. The capacity and increment size is selectable in the programming mode in sub-block F1.2. Calibration is also completed in sub-block F1.2.
Chapter 4: Maintenance and Troubleshooting Cleaning 4 Maintenance and Troubleshooting Cleaning WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING, SERVICING, CLEANING, OR REMOVING THE FUSE. FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE. Periodically clean the keyboard and covers with a soft clean cloth that has been dampened with a mild window type cleaner or detergent. DO NOT USE ANY TYPE OF INDUSTRIAL SOLVENT OR CHEMICALS. DO NOT SPRAY CLEANER DIRECTLY ONTO THE UNIT.
METTLER TOLEDO WILDCAT Scale Technical Manual Error Message Description Probable Action ---- Overload indication. Weight on scale exceeds calibrated capacity by more than 9d. Decrease load on scale. |- - - - l Underload indication. Weight on scale is below gross zero by more than 9d. Increase load on scale. | Table 4-1 Voltage Checks AC Power Test Using a Multi-meter, check the AC input power. Input power must be within -15% and +10% of the nominal AC line voltage.
Chapter 4: Maintenance and Troubleshooting Parts Replacement and Adjustments Parts Replacement and Adjustments WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE INSTALLING, SERVICING, CLEANING, OR REMOVING THE FUSE. FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY DAMAGE. CAUTION BEFORE CONNECTING OR DISCONNECTING ANY INTERNAL ELECTRONIC COMPONENTS OR INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT, ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30) SECONDS.
METTLER TOLEDO WILDCAT Scale Technical Manual 4. Reinstall the rear cover over the batteries and press into place. 5. Test for proper operation. The operating time for a new set batteries depends on how the scale is used. Estimates (without backlight) are shown below. Using the backlight feature reduces battery life. Operation Continuous use 8 hours per day Battery Life 400 hrs (Approximately 17 days) 500 hrs (Approximately 3 months) Keyboard Replacement 1.
Chapter 4: Maintenance and Troubleshooting Parts Replacement and Adjustments Load Cell Replacement 1. Remove power by removing the six “D” size batteries from the rear battery compartment and/or disconnecting the AC power adapter. 2. Remove the four screws securing the front and back halves of the cover. 3. Disconnect the keyboard tail from the Controller PCB and set the front cover aside. 4. Disconnect the load cell wiring from the J4 terminal strip on the Controller PCB. 5.
METTLER TOLEDO WILDCAT Scale Technical Manual Overload Stop Adjustment 1. The overload stops must be checked and reset if the top or bottom frame or the load cell has been replaced. 2. Remove the stainless steel platform from the base. 3. Using the proper size feeler gauge, check all six overload stops as shown in Figure 4-3. The correct gap measurements can be found in Table 4-3. 4. If the gaps are not set properly, proceed to the following steps. 5. Loosen the overload screw jam nuts.
Chapter 4: Maintenance and Troubleshooting Parts Replacement and Adjustments Shift Test A shift test verifies that the scale weighs correctly irrespective of load placement. If the scale does not pass the shift test, verify the overload stop gaps before replacing the load cell. No adjustment for the shift is possible. If the scale does not pass the shift test, the load cell must be replaced. For NTEP and Canadian Weights and Measures tests, use weights equal to 1/2 the scale’s capacity.
METTLER TOLEDO WILDCAT Scale Technical Manual Power-Up Sequence Upon power up, showing all numbers 0-9 checks the display, and a diagnostic self test is performed on the memory and microprocessor. An error message is displayed if any component fails the test. The program number [125362] is shown next, followed by the revision [Sr. 2.00]. If everything tests okay, the WILDCAT scale will show [ 0.00] on the display. The power-up sequence requires a few seconds to complete.
Chapter 5: Parts List 5 Parts List Terminal 11 8 8 22 99 3 3 44 22 55 66 2 7 7 12 12 10 10 22 11 11 Item P/N 1 2 3 127482OTC R0512000A 127483OTC 127484OTC 124466OTC 124467OTC 124454OTC 121101OTC 12901800A 14577900A 124453OTC 124464OTC 124474OTC 125967OTC 125969OTC 4 5 6 7 8 9 10 11 12 Description Front Cover, Keyboard & Overlay Assembly Screw, M4 x 6 Main PCB w/ Backlight Main PCB w/o Backlight Battery Harness Power Supply Harness Power Supply Plate Nut, Power Supply Grip Bushing Grip Bus
METTLER TOLEDO WILDCAT Scale Technical Manual Column and Base (WS30, WS60, WS60L, WS150) 1 6 2 4 7 3 5 8 Figure 5-1 Item 1 2 3 4 5 6 7 8 P/N 126431OTC 125105OTC 125103OTC 123775OTC 123976OTC 123979OTC 102504OTC 125883OTC 125881OTC 125009OTC 125006OTC 125924OTC Description Indicator Bracket Assembly Stainless Steel Column, 350 mm high Stainless Steel Column, 680 mm high Column Bracket Load Cell, 100 kg (WS30R, WS60R, WS60LR) Load Cell, 250 kg (WS150R) Load Cell Mounting Bolts, M6 × 25 Stainless Ste
Chapter 5: Parts List Column and Base (WW250 and WW500) Column and Base (WW250 and WW500) Item 1 2 3 4 5 P/N 134926OTC 6 7 131702OTC 114657OTC 118283OTC 118033OTC 102527OTC 102440OTC 136634OTC 8 9 10 126431OTC 134929OTC 102438OTC Description Stainless Steel Platter, 500 × 700mm Terminal - Refer to page 5-1 Indicator Bracket Assembly Column Assembly Wheel Mounting Bolts, M12 × 90 4” Polyurethane Wheel Set (2 wheels with brakes) Load Cell, BMI-500 kg Load Cell, BMI-1000 kg Load Cell Mounting Bolts, M1
METTLER TOLEDO WILDCAT Scale Technical Manual NOTES 5-4 (12/00)
Chapter 6: Appendices Appendix 1: Demand Mode Output 6 Appendices Appendix 1: Demand Mode Output Demand mode output occurs when a print request is received form the PRINT push button. The line format, baud rate, character length, and parity are selectable in the programming mode. Demand mode output is blocked when the scale is “in motion” or when the weight is below 20d of gross weight or over capacity. When one of those situations occurs, the print request is ignored and is not buffered.
METTLER TOLEDO WILDCAT Scale Technical Manual Single Line – Gross, Tare, Net Weights Single-line gross weight format is output if no tare weight has been taken.
Chapter 6: Appendices Appendix 1: Demand Mode Output Data Note M - - - - - L S S S P D D 9 4 Units S N P 5 4 10 C L R F 6 8 Table 6-6: Multiple Line – Net Weight Demand Format Notes 2 Gross weight legend. Selectable in the programming mode as a space, “B” for bruto or “G” for gross. 3 The gross weight data fields are padded with leading spaces to 7 digits. 4 ASCII Space character, hex value 20. 5 Weight units (lb, kg, g) as selected in setup and by the FUNCTION key.
METTLER TOLEDO WILDCAT Scale Technical Manual Appendix 2: Continuous Mode Output The continuous output format is output as the display is updated (approximately 7 times per second). The format is fixed except for baud rate, data bits, stop bits, parity, and the selectable checksum character. The continuous output mode provides compatibility with METTLER TOLEDO products that require real-time weight data (for example, Models 8624, 9323, 9325, and other accessories).
Chapter 6: Appendices Status Byte Definition The following tables detail the status bytes for the standard continuous output. Bit Identification Table For Status Byte A Bit 0 1 2 XXXX00 0 0 0 0 1 2 XXXX.XX 0 0 1 3 4 5 6 X1 1 0 Dummy Zero or Decimal Position XXXXX0 XXXXXX 1 0 0 1 0 0 Dummy Zero or Decimal Position XXX.XXX XX.XXXX 1 0 0 1 1 1 Increment Size X2 0 1 Always = 1 Always = 0 XXXXX.X 1 1 0 X.
METTLER TOLEDO WILDCAT Scale Technical Manual Appendix 3: ASCII Input Commands Beginning with software version 2.00, the WILDCAT scale has the ability to receive certain ASCII characters via the serial port to perform simple keyboard functions. Previous versions of software did not have this capability. The software version can be viewed at power up. The ASCII input commands will operate when the serial port is programmed for either demand output or continuous output.
Chapter 6: Appendices Appendix 4: GEO Codes Appendix 4: GEO Codes Use the following Geo Codes if you relocate the WILDCAT scale to a location other than the original location where it was calibrated.
METTLER TOLEDO WILDCAT Scale Technical Manual Height above sea-level in meters Northern and Southern Latitude 0 325 325 650 650 975 975 1300 2275 2600 2600 2925 2925 3250 3250 3575 0 1060 1060 2130 2130 3200 3200 4260 4260 5330 5330 6400 6400 7460 7460 8530 8530 9600 9600 10660 10660 11730 59° 32′ — 60° 49′ 24 24 23 23 22 60° 49′ — 62° 9′ 25 24 24 23 23 22 21 21 20 20 19 22 22 21 21 20 62° 9′ — 63° 30′ 25 25 24 24 20 23 23 22 22 21 21 20 63° 30′ — 64°
METTLER TOLEDO Scales & Systems 1900 Polaris Parkway Columbus, Ohio 43240 Phone (US and Canada) (800) 786-0038 (614) 438-4511 (All Others) (614) 438-4888 Internet: www.mt.com E15609900A (12/00).01 METTLER TOLEDO® is a registered trademark of Mettler-Toledo, Inc. ©2000 Mettler-Toledo, Inc.