Instruction Manual
WIKA operating instructions resistance thermometers, thermocouples
11208287.06 10/2013 GB/D/F/E
17
GB
8. Faults
7.3 Calibration, Recalibration
It is recommended that the measuring insert is recalibrated at regular intervals (resistance 
thermometers: approx. 24 months, thermocouples: approx. 12 months). This period can be 
reduced, depending on the particular application. The calibration can be carried out by the 
manufacturer, as well as on site by qualied technical sta with calibration instruments.
8. Faults
Faults Causes Measures
No signal/line breakage
Mechanical load too high or 
overtemperature
Replace probe or measuring insert 
with a suitable design
Erroneous measured values
Sensor drift caused by 
overtemperature
Replace probe or measuring insert 
with a suitable design
Sensor drift caused by chemical attack
Use a design with thermowell
Erroneous measured values 
(too low)
Entry of moisture into cable or 
measuring insert
Replace probe or measuring insert 
with a suitable design
Erroneous measured values 
and response times too long
Wrong mounting geometry, for example 
mounting depth too deep or heat 
dissipation too high
The temperature-sensitive area of the 
sensor must be inside the medium, 
and surfaces must be isolated.
Deposits on the sensor or thermowell Remove deposits
Erroneous measured values 
(of thermocouples)
Parasitic voltages (thermal voltages, 
galvanic voltage) or wrong equalisation 
line
Use a suitable equalisation line
Indication of the measured 
value jumps
Cable break in connecting cable or 
loose contact caused by mechanical 
overload
Replace probe or measuring insert 
with a suitable design, for example 
equipped with a strain relief or a 
thicker conductor cross-section
Corrosion
Composition of the medium not 
as expected or modied or wrong 
thermowell material selected
Analyse medium and then select a 
more-suitable material or replace 
thermowell regularly
Signal interference
Stray currents caused by electric elds 
or earth loops
Use of screened connecting cables, 
increase in the distance to motors and 
power lines
Earth circuits Elimination of potentials, use of 
galvanically isolated transmitter supply 
isolators or transmitters
CAUTION!
If deciencies cannot be eliminated by means of the measures listed above, shut 
down the instrument immediately, and ensure that pressure and/or signal are no 
longer present, and secure the instrument from being put back into operation 
inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions see chapter 9.2 “Return”.










