6 Commissioning, operation
80 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
6.9 Calibration Examples
Temperature calibration with the comparison method
DESCRIPTION
The user has to calibrate a temperature element (RTD or TC) in the field.
Accordingly, customer needs:
to measure the signal coming from the RTD or TC under test (DUT)
to measure the reference signal coming from the reference RTD or TC
to have a temperature generator
UP TO NOW
User has to carry in the field:
1. a standard calibrator to measure the TC or
RTD signal coming from the DUT
2. an external indicator to measure the
reference signal coming from the reference
TC or RTD
3. a dry block to generate the temperature
FROM NOW
User can use :
1. the PASCAL ET II ( 2 Input) to:
measure the TC or RTD signal coming
from the DUT
measure the reference signal coming
from the reference TC or RTD
2. a dry block to generate the temperature
ADVANTAGES
1) customer can buy only ONE CALIBRATOR, doesn’t need to buy two devices
COST SAVINGS
2) customer has to carry in the field only ONE CALIBRATOR and no more two
devices EASIER JOB
3) customer can run the calibration directly on the calibrator, storing the
calibration data FASTER CALIBRATION TIME & COST SAVINGS