Manual

6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 77
GB
X axis represents the calibration points related to the reference (REF).
Y axis represents the error in percentage (compared to the Reading range, as assigned to the
DUT) related to the Device Under Test channel (DUT).
From REPORT menu, select DEL.RPTS to cancel all the reports in the memory.
6.7 Data Logger
The LOGGER function allows to store the data displayed on the 4 Channels simultaneously.
Figure 96 Menu Logger
Pressing LOGGER, the following will be displayed:
SETUP set up of the Logger
VIEW LOG View Logs data
DEL.LOGS delete all the Logs
Selecting SETUP, “Data Logger Setup” will be displayed as per the below:
Log Name Name of the Log (compulsory)
Sampling time sampling time between 2 measurements (hours, min., sec.)
Sampling length total time of the logging
Pressing START, the logging process will start:
Figure 97 Logger data setup