6 Commissioning, operation
34 WIKA Operating Instruction, Pascal ET and Pascal ET/IS
GB
Once RTD USER is selected, it is possible to set the parameter of the RTD used.
A = X * 10
-3
°C
-1
B = X * 10
ˉ7
°C
-2
C = X * 10
-12
°C
-3
Ro = Resistance @ t = 0,01 °C
X = Value
Next step is to define the measurement range. The range selection is necessary during the
calibration procedure to define the relationship between the REF and DUT channels.
Figure 20 Range setup
Now it is possible to conclude the procedure by pressing the YES button or come back to the
previous step pressing NO.
Figure 21 Channel setup confirmation
Once the YES key is pressed, the calibrator returns to the initial page where the user can proceed
with the measurement/simulation, or assign another function to a different software channel
following the same procedure.