Instruction Manual
6 Commissioning, operation
74 WIKA Operating Instruction, Pascal 100 and Pascal 100/IS
GB
6.7 Report
To create a report one channel has to be set as REF (Reference) and another one as DUT (Device
Under Test): reference signal and direct measurement (or manual input by keyboard) are required.
For the REFERENCE: remember to set the range equivalent to the one of the DUT. When the
channel is assigned as reference, the symbol REF will appear.
For the DUT: remember to define the max error and the declaration. A bargraph will be displayed
on the DUT channel. When the channel is device under test, the symbol DUT will appear.
Press REPORT, the below menu is displayed:
VIEW RPT View Reports
SETUPRPT Set up Reports
DEL.RPTS deletes all the Reports
VIEW PRC View Procedures
SETUPPRC Set up Procedures
DEL.PRCS Delete all the Procedures
Figure 89 – Report menu
Press SETUPRPT to create a new Report.
Info required are:
Report name
DUT model (Device Under Test)
DUT Serial Number (Device Under Test)
Operator
Job Number
Found Left
Up Down
N. Points