6 Commissioning, operation
WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 37
GB
6.4.2 Other assignments
6.4.2.1 Pressure measurement
The following procedure illustrates the setting of Pascal 100 or Pascal 100/IS for pressure
measurement. The particular channel is selected by pressing one of the four big displays and then
pressing CHANNEL and ASSIGN from the menu, as shown below:
Figure 26 Channel assign selection
Automatically the program shows the following screen and it remains in this position until the
operator points out what type of parameter he desires to associate to a selected channel, for
example to the Channel 1. In this case press P for pressure.
Figure 27 Pressure parameter selection
The Pascal 100 or Pascal 100/IS can be equipped with up to two pressure modules and each of
them can handle up to three pressure sensors, two internals and one external with standard
accuracy.