Owner manual
11178035 12/2010 GB/D/F/E
Hand Spindle Pump
CPP1000-M/-L
WIKA Operating Instructions Hand Spindle Pump16
6. Mounting instruction for shut-o valve for test connec-
tion and ne adjustment valve
The mounting instruction describes the subse-
quent mounting of the shut-o valve for test
connection (Order No. 11208384) or of the ne
adjustment valve (Order No. 11248351). The
valves can be mounted at the side of the back
ange in exchange for the blind plugs which are
installed as standard.
The following combinations are possible for the individual pump versions.
CPP1000-M
Shut-o valve for test
connection
Fine adjustment valve
1 with 0
0 with 1
1 with 1
2 with 0
CPP1000-L
Shut-o valve for test
connection
Fine adjustment valve
1 with installed as standard
0 with installed as standard
If only one instrument (reference gauge or test item) with large volume is mounted on
the CPP1000-M, only one shut-o valve for test connection is needed. If both gauges
(reference and test item) have large volumes, the usage of two shut-o valves is
recommended.
When 2 shut-o valves are used, it is not possible to mount a ne adjustment valve.
GB