WIKA Operating Instructions Hand-Held Pressure Indicator22
Display Meaning Remedy
Low battery power, device will only work
for a short while longer
Replace battery
Battery flat Replace battery
Mains operation without battery: wrong
voltage
Check power supply, change it if neces-
sary
SEnS
Erro
or Err.9
No sensor connected Switch off device and connect sensor
Connected sensor or device faulty If second sensor available, check if
device is ok. Return defective device/
sensor to manufacturer for repair
Value extremely out of measuring range Check: pressure not within sensor range?
- - - -
- - - -
Logger data is being read by the serial
interface
When the data transfer is completed the
device will automatically return to normal
measuring display, no remedy necessary
No display
or strange
symbols,
device is not
responding to
input
Battery flat Replace battery
Mains operation without battery: wrong
voltage or polarity
Check power supply, change it if neces-
sary
System error Disconnect battery or power supply, wait
a while, re-connect
Instrument faulty Return to manufacturer for repair
Err.1
Measured value above allowable range Check: Is the pressure outside the
sensor‘s range? -> measured value to
high!
Sensor faulty Return to manufacturer for repair
Err.2
Measured value below allowable range Check: Is the pressure outside the
sensor‘s range? -> measured value to
low!
Sensor defective Return to manufacturer for repair
Err.3
Display range overflow Check: value above 9999 -> to high to be
displayed!
Err.4
Display range underflow Check: value below ‚-2000‘ (Tare?) ->
to low!
Err.7
System error Return to manufacturer for repair
Err.11
Value coul not be calculated Choose different unit
Calculation overflow happened Choose different unit
6. Fault and system messages