Manual
Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
GB
11221780.1.3 01/2009 GB/D/F/E
WIKA Operating Instructions Hand-Held Pressure Indicator 11
Menu Param. Setting Meaning
„Menu“
►
▲ or ▼
SEt
ConF
Set Configuration: Generic settings
Unit
mbar, bar, ... Unit: Unit of display
*, **
SL
oFF / on Sea-level correction: on or off
*, **
Alti
-200 ... +9999 Altitude: Input of altitude above sea-level [m] (only if on)
*, **
rAtE
Rate: Measuring rate (see chapter 2.3)
*
Slo Slow: Measuring rate (4 Hz filtered, low power consumption)
*
FASt Fast: Measuring rate, filtered (> 100 Hz)
*
P.dEt Peak detection: fast measuring rate, unfildered (> 100 Hz)
*
t.AVG
1 ... 120 Averaging interval in seconds, used by the averaging function off
oFF Averaging function deactivated
P.oFF
1 ... 120 Auto Power-Off delay in minutes. Device will automatically switch
itself off if, during this interval, no key is pressed, or there is no
serial communication.
oFF Auto Power-Off function inactive (continuous operation)
Out
oFF Function of the output: No output function, lowest power
consumption
SEr Output is serial interface
dAC Output is analogue output 0 ... 1 V
Adr.
01, 11 ... 91 Communicationsaddress of interface
dAC.
CH1, CH2 or
CH DIF
Choice of the input to be the source for the analogue output (if
Out = dAC)
dAC.0
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 0 V (if Out = dAC)
dAC.1
eg. -5.00 ...
+5.00 mbar
Enter desired value at which the analogue output potential should
be 1 V (if Out = dAC)
SEt
CAL
Set Calibration: Adjustment of sensors
OFS.1
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The offset of sensor 1 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF Zero displacement inactive (=0.0 °)
SCL.1
eg. -5.00 ...
+5.00 mbar
The measuring scale of sensor 1 will be changed by this factor
[%] to compensate deviations of temperature probe or measuring
device.
OFF Scale correction factor inactive (=0.000)
OFS.2
Sensordep.,
e.g. -5.00 ...
+5.00 mbar
The offset of sensor 2 will be displaced by this value to compen-
sate for deviations in the probe or in the measuring device.
OFF Zero dispacement inactive (=0.0°)
SCL.2
eg. -5.00 ...
+5.00 mbar
The measuring scale of sensor 2 will be changed by this factor
[%] to compensate deviations of temperature probe or measuring
device.
OFF Scale correction factor inactive (=0.000)