Select the sensor type by rotary switch sensor input (SW303).
Factory adjusted as K, [
].
[Table 3.2-2]
Rotary Sw.
No.
DIP Sw.
No. 7
Type of
sensor
Scale range
0OFFK -200 to 1370 °C -320 to 2500 °F
1 OFF J -200 to 1000 °C -320 to 1800 °F
2 OFF R 0 to 1760 °C 0 to 3200 °F
3 OFF B 0 to 1820 °C 0 to 3300 °F
4OFF PL- 0 to 1390 °C 0 to 2500 °F
5 OFF N 0 to 1300 °C 0 to 2300 °F
6 OFF Pt100 -199.9 to 850.0 °C -199.9 to 999.9 °F
7 OFF JPt100 -199.9 to 500.0 °C -199.9 to 900.0 °F
0 ON S 0 to 1760 °C 0 to 3200 °F
1 ON E 0 to 1000 °C 0 to 1800 °F
2 ON T -199.9 to 400.0 °C -199.9 to 750.0 °F
3 ON C (W/Re5-26) 0 to 2315 °C 0 to 4200 °F
7 ON Pt100 -200 to 850 °C -320 to 1560 °F