Electromagneticimmunity–cont
TheDREAM™serieshearingaidsareintendedforuseintheelec-
tromagnetic environment specified below The customer or the
userofaDREAM™serieshearingaidshouldassurethatitisused
insuchanenvironment
Immunity
Test
IEC 60601
Test level
Compli-
ance level
Electromagnetic environ-
ment – guidance
Portable and mobile RF
communications equip-
ment should be used no
closer to any part of the
DREAM
TM
series hearing
aid, including cables, than
the recommended separa-
tion distance calculated
from the equation applica-
ble to the frequency of the
transmitter.
Conducted
RF
IEC
61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Recommended separation
distance
d = 1.2
P
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m d = 1.2
P
80 MHz to 800 MHz
d = 2.3 P
800 MHz to 2.5 GHz