9.2 2-point correction
2-point correction can change the offset and gradient of
the input curve.
Parameter setting:
( ConF/ InP/ Corr = 2 ):
w
2-point correction is possible off-line by
means of an input signal simulator,
w
or on-line in 2 steps: correct one value first
and the second value subsequently, e.g.
after heating up the furnace..
InL: The input value of the lower scaling point is displayed. The correction function is activated via
keys ÈÌ ; the display changes from Off to the measurement value.
Adjust the lower input value by means of an input signal simulator and press key ô to confirm
the input value.
OuL: The display value of the lower scaling point is indicated.
Press keys ÈÌ to correct the lower display value and press key ô to confirm the display value.
InH: The input value of the second scaling point is displayed. Activate the corrective function by press
-
ing keys ÈÌ ; the display changes from Off to the measured value.
Adjust the upper input value by means of the input signal simulator and confirm the input value
by pressing key ô .
OuH: The display value of the upper scaling point is indicated. Correct the upper display value by pres
-
sing keys ÈÌ and press key ô to confirm the display value.
Calibrating level
44 2-point correction TB 45