Form 3785 • Price $32.00 Edition 1 • © June 1997 ONE AND TWO PEN 12 INCH CIRCULAR CHART RECORDERS AND RECORDING CONTROLLERS MRC 8000 Installation, Wiring, Operation Manual The Partlow Corporation • Two Campion Rd.
I nformation in this installation, wiring, and operation manual is subject to change without notice. One manual is provided with each instrument at the time of shipment. Extra copies are available at the price published on the front cover. Copyright © June 1997, The Partlow Corporation, all rights reserved.
TABLE OF CONTENTS SECTION 1 - GENERAL 1.1 Product Description PAGE NUMBER 5 SECTION 2 - INSTALLATION & WIRING 2.1 2.2 2.3 2.4 2.5 2.6 Installation & Wiring Unpacking Location Mounting Preparation for Wiring Wiring Connections 8 8 8 8 10 14 SECTION 3 - CONFIGURATION 3.1 3.2 3.3 3.4 3.5 Configuration (Set Up) Configuration/Jumper Positioning Operation Summary Start Up Procedures Front Panel Operation 20 21 21 22 22 SECTION 4 - OPERATION 4.1 Off Control Mode 4.2 Alarm Operation 4.
FIGURES & TABLES Figure 1-1 Figure 1-2 Figure 2-1 Figure 2-2 Figure 2-3 Figure 2-4 Figure 2-5 Figure 2-6 Figure 2-7 Figure 2-8 Figure 2-9 Figure 2-10 Figure 2-11A Figure 2-12 Figure 2-13 Figure 2-14 Figure 2-15 Figure 3-1 Figure 5-1 Table 3-1 Table 3-2 Table 3-3 Table 5-1 Table 5-2 Recorder Description Recorder Display Installation Panel Dimensions Conduit Opening Locations Noise Suppression Noise Suppression Board and Terminal Locations AC Power Input Thermocouple Inputs RTD Inputs Milliamp, Volt and Mill
Product Description 1.1 1.1.1 GENERAL The instrument is a microprocessor based circular chart Recorder or Recording Controller capable of measuring, displaying, recording and controlling from a variety of inputs. Applications include temperature, level, pressure, flow and others. The instruments can be specified as either a single or as a dual pen model. Recording, control functions, alarm settings and other parameters are easily entered via the keypad.
1.1.2 RECORDING The instrument records the selected process variable on a 12-inch circular chart. One box of standard charts is provided with each recorder. Charts are available in a wide selection of ranges. Chart rotation speed is programmable from 0.1 to 999.9 hours per revolution in 0.1 hour increments. The instrument can be ordered with one or two pens. Pen 1 is red and Pen 2 is green. Pens are the disposable fiber-tip type. 1.1.
can not be used as a control output while assigned as a process value re-transmission output. 1.1.7 DIGITAL COMMUNICATIONS The instrument can be ordered with a Digital Communications option that provides the capability of bi-directional communications with a supervisory computer. A dual pen instrument can have an individual address selected for each pen. Refer to the Communications Protocol Manual (Form 2878) for more details regarding the communications option.
Installation and Wiring 2.1 Read these instructions carefully before proceeding with installation and operation. Electrical code requirements and safety standards should be observed. Installation should be performed by qualified personnel. CAUTION: The Instrument AC power input is specified in the model number and on the wiring label affixed to the the top center of the platen. Verify the AC power input required by the instrument prior to proceeding with installation. Unpacking 2.
FIGURE 2-1 Figure 2-1A (MIN. HORZ. SPACING) 6.156" (156.36mm) 3/16" DIA. (MIN. VERT. SPACING) 3.600" (91.44mm) 14.180" (360.17mm) 12.700" (322.58mm) 10.000" (254.00mm) 0.7" (17.78mm) 12.700" (322.58mm) Figure 2-1B 2.12" (53.85mm) 12.60" (320.04mm) 17.04" (432.82mm) 2.044" (58.93mm) 5.24" (133.10mm) EC5 EC1 7.747" (196.77mm) EC6 EC2 EC3 EC4 12.600" (320.04mm) 14.12" (358.
Preparations for Wiring 2.5 2.5.1 WIRING GUIDELINES Electrical noise is a phenomenon typical of industrial environments. The following are guidelines that must be followed to minimize the effect of noise upon any instrumentation. 2.5.1.
contact with each other and amount of cross talk. Cross talk is due to the EMF (Electro Magnetic Flux) emitted by a wire as current passes through it. 2.5.1.4 USE OF SHIELDED CABLE Shielded cable helps eliminate electrical noise being induced on the wires. All analog signals should be run with shielded cable. Connection lead length should be kept as short as possible, keeping the wires protected by the shielding. The shield should be grounded at one end only.
2.5.2 SENSOR PLACEMENT (THERMOCOUPLE OR RTD) If the temperature probe is to be subjected to corrosive or abrasive conditions, it should be protected by the appropriate thermowell. The probe should be positioned to reflect true process temperature: In liquid media - the mose agitated area. In air - the best circulated area. THERMOCOUPLE LEAD RESISTANCE Thermocouple lead length can affect instrument accuracy, since the size (gauge) and the length of the wire affect lead resistance.
RTD LEAD RESISTANCE RTD lead length can affect instrument accuracy. Size (gauge) and length of the wire used affects lead length resistance. To determine the temperature error resulting from the lead length resistance, use the following equation: Terr = TLe * L where; TABLE 3 3 Wire RTD AWG No. 10 12 14 16 18 20 24 TABLE 4 AWG No. 10 12 14 16 18 20 24 TLe = value from Table 3 if 3 wire or Table 4 is 2 wire. L = length of leadwire in thousands of feet. Error °C +/-0.04 +/-0.07 +/-0.10 +/-0.16 +/-0.
Wiring Connections 2.6 All wiring connections are typically made to the instrument at the time of installation. Connections are made at the terminal boards provided, two 12 gauge wires maximum. Terminal boards are designated TB1 through TB13. See Figure 2-4 for the terminal board locations. The number of terminal boards present on the instrument depend upon the model number/ hardware configuration.
FIGURE 2-5 AC Instrument Power Input Connect the 115 VAC hot and neutral to terminals 1 and 2 respectively of TB1. See Figure 2-4 (page 14) for Terminal Board locations on the instrument. Connect the 230 VAC one leg to each terminal, be sure to check the position of the Voltage Selector switch provided with 230 VAC instruments. The switch position must match the voltage input to the instrument.
FIGURE 2-8 Volt, Millivolt and milliamp Input Make the volt, millivolt and milliamp connections as shown below. Use TB4 for thePen 1 input, and TB5 for the Pen 2 input. Terminal 1 is positive and terminal 2 is negative. The milliamp input requires the installation of an appropriate shunt resistor (ordered separately) between terminals 1 and 2. Be sure that input conditioning jumpers are in the correct positions for the input being connected. See Appendix A-1 (page 62).
FIGURE 2-10 Digital Communications Options Connections are made as shown using TB2. Refer to the Protocol Manual, Form #2878 for more details regarding the connections and how to use this option. This document is provided only when this option has been specified. If the communications network continues on to other instruments, connect the cable shields together, but not to the instrument. A terminating resistor should be installed at the terminals of the last unit in the communications loop.
FIGURE 2-12 SSR Driver Output Connections are made to relays D through A as shown. Terminal connections are made using TB7 and TB6, depending on the number of SSR Driver outputs specified. + 1 2 3 4 - SSR TB6 OR TB7 FIGURE 2-13 Current Output Connections are made to current outputs A and B as shown. Each current output is programmable as either 4 to 20 mADC or 0 to 20 mADC. Each output must be assigned to the desired function in the Program mode.
FIGURE 2-14 Transmitter Power Supply Input If the isolated 24 VDC regulated transmitter power supply has been specified, the connections should be made as shown. Connections are made using TB3, terminal 1 is positive and terminal 2 is negative. The power supply is capable of providing the power needed by up to 2 transducers (40 mADC maximum).
Configuration 3.1 After completing installation and wiring of the instrument the configuration (set up) procedures must be performed to prepare the instrument for operation on the intended application. The procedures include selecting specific parameters, entering data and possible jumper positioning. Once properly configured the instrument will retain the user selections in memory so this procedure need not be repeated unless required by changes in the application.
Manual** Stby Operation Provides for manual operation of proportional output Enable EnAb Configuration Mode security system, can lock out everything except Off and Control (See Appendix A-1, page 62, for hardware lockout information) * Applies to Recorders ** Applies to Recording Controllers Associated with each mode is a series of unique displays that are accessed via the front keypad.
Start up Procedures 3.4 All configuration parameters are listed in Tables 3-1 through 3-5. For a single pen instrument, parameters for each mode are displayed in the left display. If the instrument being configured is a two pen model, a sequence of applicable parameters will be displayed in the Pen 2 display after the Pen 1 parameters have been reviewed and configured. After the Pen 2 parameters have been completed, parameters common to both pens will be configured and displayed in the Pen 1 display.
RECORDER: ALRM1 Red ALRM2 C Red Red F Red U Red SP (Limits only) Red Green Rcdr - Lights when Alarm 1 is on. Limit - Lights when Limit has been exceeded. Lights when Alarm 2 is on. Lights to indicate that the process value is in terms of degrees C (Celsius). Lights to indicate that the process value is in terms of degrees F (Fahrenheit). Lights to indicate that the process value is in terms of Engineering units. Lights to indicate a negative displayed value.
DOWN: Used to: 1. 2. 3. 4. 5. 6. 7. Enter a mode Turn a mode Off in the Enable mode Decrease a parameter numerical value View the setpoint for Pen 2 (if provided) Decrease the setpoint value (Press hold) Step display through parameter codes in a mode Work in conjunction with other keys: a. With the SCROLL key; 1) On power up to alter the model number displayed 2) Enter Cal/Test functions b.
PROGRAM MODE FLOW CHART Prog A inPS AL2 iCor diSP * out1 * dPoS o1uL * Euu o1LL * EuL out2 * HyCo * o2uL * HyAo o2LL * rSP * KEY Actual Display AL1 rSPu * ON OFF On/Off Display Use arrow keys to turn on or off Scroll Key A * Applies to Recording Controllers Only B Numeric Display Use arrow keys to change value Up Arrow Key Down Arrow 25
B C rSPL * PoL SPuL Cru SPLL CrL AtFr * P1EC * Prnd P2EC * dFF PAEC PFF rLyA Pout rLyb Pou rLyC KEY Actual Display ON OFF On/Off Display Use arrow keys to turn on or off Scroll Key C Numeric Display Use arrow keys to change value Up Arrow Key Down Arrow 26 * Applies to Recording Controllers Only D
E D rLyd PAPu CurA Coo Com (Optional) CCon Curb CoAr Cbs Cobr CAd1 Crt CAd2 E * Applies to Recording Controllers Only 27
TABLE 3-1 PROGRAM MODE CONFIGURATION PROCEDURE Press the SCROLL key until Prog is displayed. Press the DOWN key to enter the Program mode. Pen 1 will be displayed in the left display. To enter the Pen 1 parameter, press the DOWN key. To enter the Pen 2 parameter, if provided, press the SCROLL key, then the DOWN key. To enter the unit parameter, press the SCROLL key with either Pen 1 or Pen 2 displayed until unit is displayed, then press the DOWN key.
STEP DESCRIPTION DISPLAY AVAILABLE CODES SETTINGS FACTORY SETTING YOUR SETTING 5=Current Proportioning -Direct 6=Current Proportioning -Reverse 7=Position Proportioning (open) 4* Output 1 Percent Upper Limit (o1uLand o1LL will not be seen if out1 = 0,1,2) o1uL 0 to 100 percent 100 5* Output 1 Percent Lower Limit o1LL 0 to 100 percent 0 6* Output 2 out2 0 = None (Position 0 Proportioning - Direct Closed) 1=On-Off- Direct 2=On-Off- Reverse 3=Time Proportioning- Direct 4=Time Proportioning-Re
DISPLAY AVAILABLE CODES SETTINGS FACTORY SETTING STEP DESCRIPTION 15* Hysteresis for Control Outputs HyCo 0 to 300 Width of Hysteresis Band (See page 66 for definition) 3 16 Hysteresis for Alarm Outputs HyAo 0 to 300 Width of Hysteresis Band (see page 66 for definition) 3 17* Remote Setpoint If rSP is set to zero then rSPu and rSPL are not seen rSP 0 to 2 0=Not Used 1=1 to 5 VDC 2=0 to 5 VDC 0 18* Remote Setpoint Upper Value rSPu -9999 to 9999 1400 19* Remote Setpoint Lower Valu
STEP DESCRIPTION DISPLAY AVAILABLE CODES SETTINGS FACTORY SETTING YOUR SETTING Output 2 Action on Error Condition (If out2 = 0,1,2,7 then P2EC will not be displayed) 33 Pen Action on Error Condition PAEC 0 or 1 1 0 = Pen goes to 0 % of chart 1 = Pen goes to 100 % of chart Pressing the SCROLL key with the PAEC parameter value displayed in the Pen 1 window will advance the display of a single pen instrument to the unit parameters.
STEP DESCRIPTION 44 Chart Operation in Off Mode DISPLAY AVAILABLE CODES SETTINGS Coo 0 to 1 0=Chart continues rotating in the Off mode 1=Chart stops rotating while in the Off mode FACTORY SETTING 0 COMMUNICATIONS OPTIONS PARAMETERS 32 45 Communication Configuration CCon 0 to 4 3 0=Off 1=Monitor Mode (Read Only) 2=Normal Mode (Read and Write) 3=Total Access with Limit Checking 4=Total Access without Limit Checking 46 Communication Bit Rate Selection Will not be seen if CCon=0 CbS 1 to 6 1=300
ALARM SET MODE FLOW CHART (APPLIES TO RECORDERS) Key ASEt Actual Display ON OFF PAL1 On/Off Display Use arrow keys to turn on or off Scroll Key Numeric Display Use arrow keys to change value PAL2 Up Arrow Key Down Arrow TABLE 3-3 ALARM SET MODE CONFIGURATION PROCEDURE Press and release the SCROLL key until ASEt is displayed, then press the DOWN key. Press the SCROLL key to advance the display through the parameters and their values. Use the UP and DOWN keys to select (adjust) the values.
LIMIT SETPOINT OPERATION The instrument must be in the Operate mode to allow setpoint value to be displayed and adjusted. In the Operate mode, to view the setpoint, press and release the UP or DOWN key. The displayed value is the setpoint. To change the setpoint value, press and hold the appropriate key. Press and hold the UP key to increase the setpoint or press and hold the DOWN key to decrease the setpoint. The setpoint will change slowly at first then faster as the key is held.
TABLE 3-2 TUNE MODE CONFIGURATION PROCEDURE (APPLIES TO RECORDING CONTROLLERS) The Tune mode allows the entry, review or altering of the process control Tune adjustments and alarm setting(s). To enter the Tune mode, press and release the SCROLL key until tunE is displayed, then press the DOWN key. Press the SCROLL key to advance the display through the parameters and their values. Use the UP and DOWN keys to select (adjust) the values.
DISPLAY AVAILABLE CODES SETTINGS FACTORY SETTING STEP DESCRIPTION 11 Automatic Reset Integration (Will be seen if Pb1 or Pb2 was shown) ArSt 0.0 to 100.0 repeats per minute 0.0 12 Rate Derivative (Wll be seen if Pb1 or Pb2 was seen) rAtE 0.0 to 10.0 minutes 0.
TABLE 3-3 ENABLE MODE CONFIGURATION PROCEDURE To enter the Enable mode, press the UP and DOWN keys while in OPEr, CtrL, or oFF modes. All the display lamps will light. After 2 seconds, the display will show Cchg and the pen(s) will move to and remain at a point above the top graduation on the chart. Continue to press the UP and DOWN keys, after 2 additional seconds, the display will show P dn and the Pen(s) are driven below the bottom graduation on the chart.
ENABLE MODE FLOW CHART RECORDERS EnAb RECORDING CONTROLLERS Press UP and DOWN ARROWS for 10 seconds to enter this loop.
Operation 4.1 4.1.1 OFF MODE In the Off mode, the instrument control and alarm function (s) are turned off. Process Retransmission signal(s) remain active. The chart rotation can be selected in the Program mode to stop or continue to rotate when the instrument is in the Off mode. The pen(s) will remain active. The Off mode can be entered by pressing and releasing the SCROLL key until the display reads oFF, then pressing the DOWN key.
Dual Pen Instruments: If the instrument is specified and provided with the Remote Setpoint capability for either or both pens and the Remote Setpoint has been properly configured in the Program mode for either or both pens, the Setpoint Select mode will be accessible, if enabled. Press and release the SCROLL keys until SPS appears in the display, then press the DOWN key. The left display will light and the upper display will blank if pen 1 is selected for a remote setpoint.
4.1.2.4 TIME PROPORTIONING CONTROL Time Proportioning Control can be implemented on controllers provided with SPST relay or SSR driver output(s). Time proportioning can be programed for output 1 and/or 2 for each pen. Time proportioning control is accomplished by cycling the output On and Off when the process value is within the proportional bandwidth selected at a prescribed time period. The time period is selected in the Tune mode by adjusting Ct1 and/or Ct2. The On time is a percentage of the Cycle Time.
4.1.2.7 PROPORTIONAL OUTPUT PERCENTAGE DISPLAY While in the Control mode, pressing the UP and the SCROLL keys at the same time will cause the display to sequence through a series of display codes and values: Po1 Po2 Proc Percent Output (if applicable) Percent Output (if applicable) Process Value Output 1% value Output 2% value Actual Process Value Each code and output value will be displayed only if the corresponding proportional output is present. Each code or value will be displayed for 1 second.
STANDBY MODE FLOW CHART (APPLIES TO RECORDING CONTROLLERS) Stby KEY Actual Display Po1 ON OFF On/Off Display Use arrow keys to turn on or off Scroll Key Po2 Numeric Display Use arrow keys to change value Up Arrow Key Down Arrow Alarm Operation 4.2 There are two alarms available per pen. The type of alarm is selected in the Program mode as follows: 1. Process Alarm Direct - the alarm will be On if the process value is greater than the process value selected. 2.
Tune Mode Operation 4.3 (Applies to Recording Controllers) Proportional output controllers may require the adjustment (tuning) of the PID and other related parameters. This provides a means for the instrument's control algorithm to be adjusted to meet specific application requirements. 4.3.1 SYSTEMATIC TUNING METHOD 1. Changes in tuning parameters should be made one at a time. 2.
parameters over a period of time to obtain optimal control of the process. 4.3.2 ZIEGLER NICHOLS TUNING METHOD This procedure has been determined empirically to yield ideal 1/4 amptitude decay tuning parameters that are determined by watching the system in a sustained oscillation (curve C, page 44, the ultimate proportional band and ultimate time period) and then using these values from this sustained oscillation to calculate ideal parameters.
Period C B A Curve A : unstable Curve B : stable Curve C : continuously cycling, ultimate PB and period 46
Service 5.1 This section contains information regarding calibration and test procedures that can be performed in the field as well as items concerning the normal maintenance of the instrument. Changing Charts 5.2 Chart changes may be done while in the normal operating mode. CAUTION: The chart flange assembly pin is sharp to perforate the chart. Use caution while installing the chart to avoid coming into contact with the pin. 1. Depress and hold the UP and DOWN keys for between 2 and 3 seconds.
Calibration 5.4 CAUTION: Do not attempt any calibrations without the proper test equipment that meets or exceeds the specifications listed. Press and release the SCROLL key until CAL appears on the display , then press the DOWN key to enter the mode. The display will change to CAL1. Press the SCROLL key to advance the display to the other calibration modes available. For two pen units, CAL5 will only need to be required on TB4 to calibrate Pen 1 and Pen 2 inputs.
TABLE 5-1 CALIBRATION PROCEDURES Calibration Procedure CAL 1 Description Reinitialization of program and tuning values. CAL 2 Main calibration necessary for all input types. CAL 3 Cold Junction Compensation calibration used to correct for component variation in the CJC circuit. Necessary for thermocouple inputs. CAL 4 Cold Junction Utility, displays temperature the cold junction compensator is sensing. No adjustment is made with this procedure. CAL 5 RTD input calibration.
Error recovery: See section 5.6 (page 55) for details. Ensure that the millivolt source is connected correctly and functioning properly. The calibration can be exited when hLd1 or the calibration reference number is displayed by pressing the SCROLL key. CAL2 QUICK CALIBRATION This routine will allow the operator to execute a rough calibration on their unit via the keypad with no other equipment or disturbance to established wiring.
5.4.5 CAL 5 RTD INPUT This procedure determines and saves calibration values relating to RTD inputs. This calibration must be preceded by CAL 2 to properly calibrate the instrument. Both RTD inputs must be calibrated and both inputs must have valid inputs during the calibration. Decade boxes with .01% resolution or equivalent are required. Make sure that the Processor board jumpers JU4, JU6 and JU5, JU7 are in the proper positions. See Appendix A-1 (page 62).
5.4.7 CAL 9 PEN CALIBRATION This procedure is used to calibrate the pen(s). No special test equipment required. Valid inputs must be connected to TB 4 and TB 5 before performing this calibration. With CAL 9 displayed, push and hold the DOWN key, then press the SCROLL key . Release both keys and the display will indicate PEn1. For 2 Pen instruments, press the DOWN key to toggle the display between pen 1 and pen 2. With desired pen displayed, press the SCROLL key. P.
Test Mode 5.5 To enter the Test mode, press and release the SCROLL key until tESt appears on the display then press the DOWN key. tSt1 will be displayed, press and release the SCROLL key to advance the display to the desired test. Tests 1, 2 and 3 are performed as a unit so the display will advance directly to tSt4 from tSt1. Listed in Table 5-2, page 52, are the test procedures available. Test 1, 2 and 3 are performed on start up, periodically during operation, and on entry into the Test mode.
TABLE 5-2 TEST PROCEDURES AND DESCRIPTION Test Test 1 Description Microprocessor internal RAM test. Used to check the processor RAM to make sure it is functioning correctly. Test 2 External RAM test, used to test the RAM chip for proper function. Test 3 EPROM checksum test, used to check that the EPROM program is correct. Test 4 External RAM checksum test; instrument test and identifies how many times Errors16 or 17 have occurred.
Test 4 can be used to display the line frequency determined at power-up. With tSt4 displayed, press and hold the DOWN key and then press the UP key. The display will blank momentarily, then show the line frequency it measured, then blank momentarily, and then display tSt4. The frequency measured, and displayed may not be exactly 50 or 60, but the software will assume it is 50 or 60 during normal operation. 5.5.
terminal 2 is negative. Press the SCROLL key and the display will indicate 4 milliamps. Use the UP and/or DOWN key to increase or decrease the current output in 1 mADC steps from 0 to 20 mADC. The current output reading should be ± 0.1 mADC at any output value. A ± 5 % of span adjustment for the current output(s) is provided by using the potentiometer adjacent to the current output on the Current Output board. See Appendix A-4 (page 65). To exit the test, press the SCROLL key and tSt7 will be displayed.
Send mode. When the sending instrument displays rEC, the receiving instruments should display rEC1. To terminate the test, press the SCROLL key for one second. Upon exit, tStA will be displayed. The existence of the communications hardware is dependent on the hardware configuration. Trouble-shooting and Diagnostics 5.6 The Trouble-shooting Guidelines Section consists of two columns. The first column is a list of some possible instrument conditions.
Model Number Displayed during power up is incorrect 1. Turn off the instrument power, wait 5 seconds then re-apply the power. Verify that the number displayed during the power up sequence is the same as indicated on the label affixed to the platen. If the number displayed is incorrect, perform the following steps: a. Turn off the power to the instrument. Press and hold the UP and DOWN keys.
5. If the output appears not to turn off remove the power to the instrument. Open the cover and loosen the platen hold down screw. Swing the platen open. Clip the resistor located on the Relay Board adjacent to the output(s) that seem to stay on (See Appendix A-2, page 63). A .01 microfarrad, 1 KV capacitor should be connected from the terminal listed below, for the output where the resistor indicated was removed, to the AC ground.
ERROR CODE DISPLAYED 60 SnSr - Sensor Break or Over Range 1. Inspect the sensor for proper operation and connection to the instrument. Acceptable sensor ranges for the instrument are listed in the Specifications section of Appendix D (page 69). 2. Verify that the Program mode input selection matches the sensor input connected (page 28 or Software Ref. Sheet, page 72, if already filled out). 3.
Er 4 - RTD Mismatch Error 1. Check the connections to the instrument for the RTD Input Calibration (CAL5) as described in the Calibration section (page 49). Repeat the RTD Input Calibration. Er 5 - No Zero Crossings Detected 1. Turn off the power to the instrument. Wait 5 seconds, and turn the power on. 2. Connect the instrument to another AC power source. Er 6 - AC line below 43 HZ 1. Turn off the power to the instrument. Wait 5 seconds, and turn the power on. 2.
Er13 - RTD CAL5 Input Error 1. Check that resistance device is of the correct value and properly connected to the instrument and is within the tolerance limits as indicated in the CAL5 procedure of the Calibration section (page 49). 2. Turn off the power to the instrument and open the cover. Loosen the platen hold down screw and swing the platen open.
Momentary Er 72 Incorrect Block Check character was received 1. Data received not valid, possible corruption on the comm line. Possible noise problems. Check baud rates, instrument addresses, line connections, and termination. Momentary Er 73 Byte received with incorrect parity. 1. Check baud rate. 2. Possible noise problems. Check instrument addresses, line connections, and termination.
Appendix A Board Layouts FIGURE A-1 - PROCESSOR BOARD Top SWI for Rev. Y and above UNLOCKED 230 JU1 ENABLE MODE SWI for Rev.
FIGURE A-2 - SPST RELAY/SSR DRIVER OUTPUT BOARD TB6 TS1 R1 TB7 TS2 R2 TS3 R3 TS4 R4 C1 C2 C3 C4 K1 K2 K3 K4 (rlyA) (rlyB) (rlyC) (rlyD) P7 If the relay is connected to a high impedance AC device, the snubber network used to protect the relay contact may cause the output to appear to be activated when the relay is off. To cure the problem, cut the snubber resistor for the output that is being affected.
FIGURE A-3 - CURRENT OUTPUT BOARD R27 R7 TB10 TB11 If this option board was ordered, you will find it located in the lower right hand corner of the instrument.
Appendix B Glossary Automatic Reset (Integral) This parameter is used so that the instrument will compensate for process variable deviations from setpoint that occur when the process load characteristics change. Instructions for determining the automatic reset settings are given in Table 3-2 (Page 34). Factory default is.0.0. Display code ArSt.
Pen Action on Power Up This parameter specifies whether the pen, on a power-up will drive to the "Home Position" (center of chart), then return to its correct postion. This is done as a cal check. Settings are 0=go to "home" and 1=remain in last position prior to power down. Default is 0. Platen The flat surface in the instrument upon which the chart rotates. Position Proportioning Sensitivity A percentage of the first output proportional band width (Pb1).
Appendix C Order Matrix 8 1 2 PEN 1 Recorder Only Recording Controller 0 1 2 PEN 2 None Recorder Only Recording Controller 0 1 2 4 *RELAY OUTPUTS None One SPST Two SPST Four SPST 0 1 2 4 *SSR DRIVER OUTPUTS None One Two Four 0 1 2 4 TO 20mA OUTPUTS None One Two 0 1 TRANSMITTER POWER SUPPLY None 24VDC Regulated/Isolated 0 1 2 PEN 1 AUXILLIARY INPUT None Position Proportioning Remote Setpoint 0 1 2 PEN 2 AUXILLIARY INPUT None Position Proportioning Remote Setpoint 0 2 DIGITAL COMMUNICATIONS N
Appendix D Product Specifications INPUTS Input Types/Range Thermocouple Type Range J 0°C to 760°C K 0°C to 1360°C T -200°C to 400°C R 200°C to 1650°C S 200°C to 1650°C E 0°C to 750°C B 200°C to 1800°C N 0°C to 1300°C C 200°C to 2300°C 100 ohm Platinum -140°C to 400°C .00385 ohms/ohm/°C 0 to 20mA, 4 to 20mA External Shunt Resistor, 2.5 ohms or 250 ohms 0 to 25mV, 0 to 50mV, 10 to 50mV, 0 to 5 V, 1 to 5 V > 100M ohm for TC and mV inputs 100K ohms for 5V inputs 2.
RECORDING Pen Type Pen Color Chart Size Chart Drive Chart Rotation Chart Span Filtering Disposable fiber tip Pen 1 - Red Pen 2 - Green 12 inch Stepper motor User configurable: 0.1 to 999 hours, in 0.1 hour increments Bottom and top of span, -9999 to 9999 units 1 to 20 scans averaging (affects display and recording) RECORDING PERFORMANCE Chart Recording Accuracy Resolution Dead Band Response Time Chart Rotation Accuracy 0.5% of chart span reference accuracy 0.15% of chart span 0.
POWER REQUIREMENTS Line Voltage Power Consumption: 115/230 VAC, ±10%, 50/60 Hz 25 VA Maximum CONSTRUCTION Enclosure NEMA Rating Conduit Openings Mounting Overall Dimensions Panel Cutout Panel Depth Panel Protrusion Weight Gasketed cover, case, and windows. Structural foam case and coved with plastic or glass window area. Door lock available NEMA 3 standard, NEMA 4X optional Four openings standard, 2 additional as required Panel or wall or optional pipe mounting 14.12" wide X 16.77" high X 7.
Appendix E Software Reference/Record Sheet Recorders: Pen 1 ENABLE MODE EtSt ECAL EPro EASt ESPC ALARM SET Pen 1 PAL1 PAL2 Pen 2 inPS iCor AL1 AL2 disP dPoS Euu EuL HyAo SPuL SPLL Prnd dFF PFF Pout Pou PoL Cru CrL PAEC PROGRAM MODE Pen 2 unit rLyA rLyb rLyC rLyd CurA Curb CoAr Cobr Crt PAPu Coo CCon CbS CAd1 CAd2 73
Recording Controllers: Pen 1 inPS iCor out1 o1uL o1LL out2 o2uL o2LL AL1 AL2 diSP dPoS Euu EuL HyCo HyAo rSP rSPu rSPL SPuL SPLL AtFr Prnd dFF PFF Pout Pou PoL Cru CrL P1EC P2EC PAEC 74 PROGRAM MODE Pen 2 rLyA rLyb rLyC rLyd CurA Curb CoAr Cobr Crt PAPu Coo CCon CbS CAd1 CAd2 unit TUNE MODE Pen 1 Pen 2 SoP PAL1 dAL1 bAL1 PAL2 dAL2 bAL2 Pb1 Pb2 rSEt ArSt rAtE Ct1 Ct2 SEnS FoP ENABLE MODE EtSt ECAL EPro Etun ESby ESPS ESPC
Warranty and Return Statement These products are sold by The Partlow Corporation (Partlow) under the warranties set forth in the following paragraphs. Such warranties are extended only with respect to a purchase of these products, as new merchandise, directly from Partlow or from a Partlow distributor, representative or reseller, and are extended only to the first buyer thereof who purchases them other than for the purpose of resale.
ISO 9002 Certified PARTLOW CORPORATION • 2 CAMPION ROAD • NEW HARTFORD, NY 13413 USA • 1-800-866-6659 • 315-797-2222 • FAX 315-797-0403