Manual
29
6 Decimal Position dPoS 0=None
0
1=One decimal position
2=Two decimal positions
3=Three decimal positions
RTD and thermocouple inputs
are limited to either 0 or
1 decimal positions
7 Engineering Units Euu -9999 to 9999
1000
Upper Value (If Input
Select = 30, 31, 32, 33, 34)
8 Engineering Units EuL -9999 to 9999
0
Lower Value (If Input
Select = 30, 31, 32, 33, 34)
9 Hysteresis for HyAo 0 to 300 degrees/units
3
Alarm Outputs Width of Hysteresis Band
(See page 54 for definition)
10 Setpoint Upper Limit SPuL -9999 to 9999
1400
(Limit Only)
11 Setpoint Lower Limit SPLL -9999 to 9999
0
(Limit Only)
12 Process Rounding of Prnd 0=No Rounding
0
Displayed value 1 to 100 degrees/units
13 Display Filter Factor dFF 1=No filtering
1
1 to 20 units
(number of values
averaged)
14 Process Filter Factor PFF 1 = No filtering
1
1 to 20 units
(number of values averaged)
15 Process Value Output Pout 0=Not selected
0
(If Pout =0 then Pou and 1=Selected
PoL will not be displayed)
16 Process Output Pou -9999 to 9999 degrees/units
2000
Upper Value
17 Process Output PoL -9999 to 9999 degrees/units
0
Lower Value
18 Chart Range Upper Value Cru -9999 to 9999 degrees/units
200
19 Chart Range Lower Value CrL -9999 to 9999 degrees/units
0
20 Pen Action on PAEC 0=Pen goes to 0%
1
Error Condition of chart span
1=Pen goes to 100%
of chart span
Pressing the SCROLL key with the PAEC parameter value displayed in the Pen 1 window will
advance the display of a single pen instrument to the unit parameters. Pressing the SCROLL
key with the PAEC parameter displayed in the Pen 1 window of a two pen instrument will
advance the display to be inPS in the Pen 2 window. The Pen 2 Program mode parameter
selections can be made now. Pressing the SCROLL key with the PAEC parameter value
displayed in the Pen 2 window will cause the display to advance to the unit parameters.
DISPLAY AVAILABLE FACTORY YOUR
STEP DESCRIPTION CODES SETTINGS SETTING SETTING