Filter Definitions (Infinite F200, F200 Pro, F500)
Select the appropriate filter position and enter the new wavelength, bandwidth,
and measurement mode for each new filter:
Measurement
Mode:
Choose from the dropdown list ‘FI’ for fluorescence
intensity, ‘ABS’ for absorbance measurements, FP for
fluorescence polarization and ‘Empty’ for filter-free
positions.
Wavelength:
Enter the filter wavelength. For fluorescence intensity and
fluorescence polarization measurements, set the filter
wavelength within the allowed range of the connected
instrument. Absorbance filters are definable between 230
and 1000 nm (Excitation only).
Bandwidth:
Enter the bandwidth (nm) of the filter.
Description:
This field can be used for individual user remarks about the
filter, e.g. filter name, application, etc.
Purchase Date:
This option enables the user to enter the purchase or
installation date of the filter.
Flash Counter:
The flash counter monitors the number of flashes through
a filter. The flash counter number provides the user only
with additional information about the filter in use.
For a new filter, set the counter to 0. For a previously used
filter, enter the last collected flash number if the number is
available. The flash counter number is saved together with
other information about the filter on the filter slide
microchip. If you replace a filter, this information will be lost
unless the last filter flash number is manually documented
by the user.
Confirm the new filter values by clicking Save. Close the Filter Definition dialog
and the system is ready to perform measurements with the new filters.
Refer to the Instructions of Use of the connected instrument for further details and
examples.