Specifications

3. Defining Measurements
2011-03 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 55
3. Defining Measurements
The following chapter describes some examples to illustrate the definition of
different measurements.
The Infinite M1000 instrument offers the Quick-Start-Script button in the front
right corner on the top cover of the instrument. It may be used to start favorite
measurement scripts directly from the instrument.
3.1 Defining End Point Measurements
The following example describes an Absorbance End Point Measurement in all
wells of a 96 well plate:
1. Select a 96 well plate (e.g. Greiner 96 Flat Transparent) from the Plate
definition drop-down list. If the Part of Plate program element is not
visible, click the link
Use a part of the plate
2. Double-click the Absorbance program element from the Control bar,
and define the Workflow as follows:
. It is recommended to use the
Part of Plate program element in every workflow, even if all wells are
measured.
3. Wavelength/Measurement: 492 nm
4. Read/Number of reads/flashes: 25 (per well)
5. Settle time (time between moving the plate and starting the
measurement): 0 ms: