Specifications
2. Measurement Parameter Editor
2011-03 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 51
Enter the respective parameters:
Timer Enter the Wait time (hh:mm:ss)
Options Wait for injection: The time for injection is included in the
wait time.
Ignore wait at last kinetic cycle: When the program step
Wait (Timer) is the last action within a kinetic run, the wait
time will be ignored in the last cycle.
Incubation
Incubation is always done at the heating position to ensure inside the instrument
proper temperature distribution.
Incubation can consist of shaking and waiting steps (up to 2 shaking steps and up
to 2 waiting steps are allowed in any combination).
The Remaining Wait step waits until the overall incubation time is over (including
shaking and waiting times).
The incubation program element is typically used to perform shaking and waiting
at a certain temperature for a certain time.
The incubation stripe of the Infinite F50 contains only an input field for the
Incubation time.
Example for the Infinite F50
Enter the appropriate parameters for incubation:
Incubation time
Enter the total time (min. 5 s)
Actions Available actions: Shaking, Wait (Timer)
2 wait and 2 shaking actions are allowed.
Select actions by double-clicking or use the arrow keys.
Organize actions by using the up/down keys.
Remaining Wait (Timer): mandatory, cannot be deleted
or edited (duration 3 s)