Specifications

2. Measurement Parameter Editor
2011-03 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 49
After having started the measurement, it is possible to interrupt a plate-wise
kinetic measurement clicking the Pause button and to continue:
Kinetic Condition
Use the Kinetic Condition program element to define which actions should be
executed at a certain cycle.
If 3 is entered for Execute command at cycle within a kinetic measurement
containing, e.g. a Shake step, shaking is performed only at cycle 3.
Note
Kinetic conditions such as Shake, Inject and Dispense should be inserted
right after a Kinetic Cycle program element in order to ensure optimal
result reproducibility.
Users are advised to set up suitable scripts prior to the measurements
and to use the same script for all similar kinetic measurements in order
to obtain comparable results.