Specifications

2. Measurement Parameter Editor
48 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 2011-03
2.2.4 Kinetic
Kinetic Cycle
Use the program element Kinetic Cycle to perform several consecutive
measurements, which may be executed in certain intervals.
Enter the respective parameters:
Cycles Number of cycles: Enter a number or click the up or down
arrows for the number of actual measurement steps (2 1000
cycles)
Duration: Enter the duration, format hh:mm:ss.
Kinetic
Interval
Use kinetic interval: Enter the time interval (hh:mm:ss or
ms).
Plate-wise kinetic measurements
Each cycle of the kinetic measurement is performed on all selected wells. Plate-
wise kinetic measurements may contain a maximum of ten independent
measurement stripes that do not need to be of the same measurement type.
Well-wise kinetic measurements
All cycles of the kinetic measurement are first performed in one well before
continuing to the next well. Well-wise kinetic measurements may be composed of
a maximum of four measurement stripes of the same type, e.g., four absorbance
stripes. The Infinite M1000 allows five measurement stripes of the same type
within well-wise kinetic measurements.