Specifications

2. Measurement Parameter Editor
2011-03 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 43
Wavelengths From: Select the starting wavelength for the scan.
To: Select the endpoint wavelength for the scan.
Step: enter a valid value
Bandwidth: Select a value from the drop down list.
Mode Select Top or Bottom.
Gain Default: this value is instrument specific (see also
30036266_IFU_InfiniteM1000)
Manual: User-defined gain value (valid range 1-255)
Integration Integration time: enter a value to specify the
duration of integration.
Z-Position Default: 22000 µm
Manual: The Z-position represents the height of the
measurement head above the microplate.
Label
Type in a label name.
2.2.3 Actions
Temperature
Select the Temperature program element to enter a certain target temperature.
Select On to enter a target temperature value. Click on the link Wait until
temperature is reached
The measurement only starts if the current instrument temperature is within the
specified range. See
to define the Minimum and/or Maximum temperature
values. The heating of the instrument starts when clicking the Start button. For
pre-heating the instrument, select Heating… in the Instrument menu and click
the On button.
2.2.5 Miscellaneous/Wait for Temperature.