Specifications

2. Measurement Parameter Editor
42 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 2011-03
Filters
Use of Color Filters for Single Luminescence:
(available for Infinite F500 and M1000)
All filters that are available for dual color luminescence may be
used in single luminescence measurements as well. Besides
the attenuation functions an additional dropdown list in the
attenuation field displays the filters for GREEN, GREEN1,
BLUE and MAGENTA to be selected individually for single
luminescence applications.
Integration
time
Enter a value to specify the duration of integration. All wells will
be measured with this fixed user-defined integration time.
Settle time
Enter a value to specify the time delay between a plate
transport movement and the start of integration.
Luminescence Dual Color
The Luminescence Dual Color program element is used to discriminate
different wavelengths within the luminescence signal (for assays that are based
on 2 distinct signals).
This dual filter system permits independent measurement by detecting two
different wavelengths within one well.
The following are the Fluorescence Dual Color parameters:
Parameter Select the filters Green, Green1 and Blue1 or Magenta, and
enter the Integration time for each label.
If required, enter a Settle time before the measurement.
Label Enter different Label Names.
Luminescence Scan
The Luminescence Scan program element is available with the Infinite M1000
with main firmware V 2.0 or higher (Ref 30061442 ).