Specifications

2. Measurement Parameter Editor
2011-03 Instructions for Use for i-control No. 30018668 Rev. No. 1.8 39
Multiple Reads per Well
The i-control software allows the user to define multiple reads per well (MRW) in
Absorbance, Fluorescence top and Fluorescence bottom mode.
The MRW feature is not available for well wise measurements.
The Reference wavelength on the absorbance program element is not
selectable in combination with multiple reads per well.
For Infinite M1000 instruments there is a minimum Settle time of 10 msec
required as soon as Multiple reads per well is selected.
The multiple reads per well function can be activated on an absorbance or
fluorescence intensity program element by selecting the Multiple reads per well
check box:
Note
The function Multiple reads per well is only available for the fixed
wavelength reading modes Absorbance, Fluorescence intensity top and
Fluorescence intensity bottom. The function is not available for scan
measurements.
More details on defining parameters for multiple reads per well, are available in
the respective Instructions for Use of the instrument connected.
The multiple reads per well function is available for plate formats with up to 384
wells. 1536 well plates are not supported.