Specifications

Transmitter M200 easy 39
© 09 / 09 Mettler-Toledo AG, CH-8606 Greifensee, Switzerland Transmitter Multiparameter M200 easy
Printed in Switzerland 52 121 501
Enter the Process Pressure. The units for ProcPres and CalPres do not have to be the same.
For the algorithm of the process calibration the applied pressure (ProcCalPres) has to be
defined. The value of the process pressure (ProcPres) or the calibration pressure (CalPres)
can be used. Chose the pressure, that applies during the process calibration, resp. should be
used for the algorithm and press [ENTER].
The salinity of the measured solution and the relative humidity of the calibration gas can also be
entered. The allowed values for Relative Humidity are in the range 0% to 100%.
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard
the entered values and return to the measurement display screen, selecting Yes will save
changes made.
8.2.4 Set Averaging
Press the [ENTER] key to select this Menu. The averaging method (noise filter) for each
measurement line can now be selected. The options are Special (Default), None, Low, Medium
and High:
None = no averaging or filtering
Low = equivalent to a 3 point moving average
Medium = equivalent to a 6 point moving average
High = equivalent to a 10 point moving average
Special = averaging depending on signal change (normally High averaging but Low averaging
for large changes in input signal)
Pressing the [ENTER] key again will bring up the Save Changes dialog. Selecting No will discard
the entered values and return to the measurement display screen, selecting Yes will save
changes made.