20 Setting up the Vision Screener Welch Allyn Spot Vision Screener
Import
With the import function, you can import:
• Updated criteria files (SpotCriteria.csv)
• Future, queued, screening files (SpotSubjects.csv or Location.csv By using a
Location in the name of the file, this will default all subjects in the file to a
location via the name of the file, ex LakeMary.csv would default to
“LakeMary” as the location for all subjects in this file)
• Custom banner for the screening reports (banner.png) The file must follow
the correct naming convention for a successful import (i.e., banner.png,
SpotCriteria.csv, SpotSubjects.csv. Remember SpotSubjects can have a
location as the name of the file).
• Updated criteria files (SpotCriteria.csv)
Custom Banner
Welch Allyn offers the ability to add a custom banner to the bottom of the printable
reports. Please note that Welch Allyn does not supply custom banners. You will need to
create your banner using the specifics below:
• Banner files must be named “banner.png”
• No larger than 1MB
• Image must be 1376 x 240 pixels
To add a custom banner
1. Place the banner.png file on the root/top-level directory of the USB.
2. Insert the USB in the vision screener. Go to To o l s , then select Import/Export.
3. The import file option will highlight; select Import.
4. To verify that the file was imported successfully, screen a new subject or select a
completed record and print the report. Your new banner will be printed on the bottom
of the report.