Setting up the Vision Screener
Charging
1. Connect the provided DC cord set and power supply/charger together.
2. Gently lift the back of the vision screener to view the available connections.
3. Locate the DC power connector on the vision screener and connect the power
supply/charger. The DC power connector on the power supply/charger will easily
insert into the DC power connector on the vision screener. Using force to connect the
power charger can damage the device and will void the warranty.
4. Plug the DC power cord into an available DC wall outlet to charge the device.
Powering the vision screener on and off
1. To turn the vision screener on, press and release the power button (start-up process
takes approximately 30 seconds). Once powered on, you will see the Main Menu
2. To turn the device off, press and hold power button for two seconds. A confirmation
screen will appear to confirm shut down.
3. A flashing green light means the unit is charging, a steady green light means the unit
is fully charged.
Tripod Mount
The vision screener can be mounted to a standard photographic tripod. A ¼-20 threaded
mount located on the bottom of the device is available for attaching a tripod.
Helpful Hints
• No activity on the vision screener—After 60 seconds the screen will dim; after 5
minutes the screen will automatically turn black to conserve battery power. Tap the
screen to ‘wake up’ the vision screener or press the power button. If the screen does
not light up, plug in the power cord and try again.
• Exporting/delete historical records—Verify you have successfully exported desired
data to a USB drive before you select Delete All Records.
• Routinely delete ALL
records from the vision screener—Go to the History screen
and select Clear All icon. Select Delete All Records. You may also delete all queued
records.
• Avoid powering off during a screening—Once the screening takes place, exit by
selecting the Exit button, then power down the device normally.