Specifications

110 Reference Welch Allyn Propaq CS Vital Signs Monitor
Monitor (physical) specifications
Characteristic Specification
Protection classifications, all configurations
a
a. Per EN 60601-1 unless otherwise stated.
Type of Protection against Electric
Shock—Power Adapter
Power adapter class 1
Type of Protection against Electric
Shock—Monitor (connected to power
adapter or internal battery)
Protective earth not available in monitor. Monitor designed and tested to meet
Double Insulation Requirement.
Degree of Protection Against Electric
Shock, for Parts Applied to Patients
See monitor labels
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Electrosurgery interference
suppression
Suitable for use in the presence of electrosurgery
Method of Disinfection Not suitable for autoclaving (see cleaning instructions, “Inspect and clean the
monitor and accessories” on page 86)
Flammable Anesthetics Not suitable for use with flammable anesthetics
Monitor only
Height 8.2 in (20.8 cm) with handle
Width 9.6 in (24.4 cm)
Depth 5.6 in (14.1 cm)
Weight 7.6 lb (3.4 kg)
Monitor with SpO
2
module
Height 8.2 in (20.8 cm) with handle
Width 9.6 in (24.4 cm)
Depth 7.7 in (19.7 cm)
Weight 10.8 lb (4.9 kg)
Monitor with expansion module
(Printer / SpO
2
/ MCO
2
)
Height 11.4 in (28.8 cm)
Width 9.6 in (24.4 cm)
Depth 7.7 in (19.7 cm) with back feet
Weight with Printer, SpO
2
, and MCO
2
14.4 lb (6.5 kg)