Specifications
Directions for Use Reference 109
Wireless Propaq CS Radio specifications (2.4 GHz)
Monitor (environmental) specifications
Characteristic Specification
FlexNet Network 2.4 GHz Wireless Local Area Network (WLAN) and 10/100 Base-T Ethernet network
Frequency
a
a. When used within certain countries, authorization for use is restricted as follows:
France The equipment is internally restricted to the 2.448-2.482 GHz frequency range.
Spain The equipment is internally restricted to the 2.447-2.473 GHz frequency range.
Japan The equipment is internally restricted to the 2.473-2.495 GHz frequency range.
Italy Operation requires a user license.
Note: The frequency ranges specified above are subject to geographic-specific regulatory authorities.
2.402 to 2.480 GHz
Modulation GFSK, Frequency Hopping Spread Spectrum (FHSS)
Output Power 112 mW (maximum)
IEEE 802.11 compliant Yes
Wireless Propaq CS monitors
per Access Point
10 (max.)
Characteristic Specification
Operating Temperature 0° to 40° C
Shipping and Storage Temperature -20° to 60° C
Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m)
Shipping and Storage Altitude -2,000 to 40,000 ft (-610 to 12,192 m)
Operating Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shipping and Storage Relative Humidity 15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural
Shock 50 g
Vibration, Random 0.02 g
2
/Hz from 10 to 500 Hz, ramping down to 0.002 g
2
/Hz at 2000 Hz.
Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA DO-
160D, Category C.
Degree of protection against ingress for
monitors without CO
2
or printer options
IPX1 rating, drip proof per EN60529: 1991
Caution The monitor may not meet performance specifications if it is not used
or stored within these environmental specifications.