Specifications

24 Understanding the LCD messages Welch Allyn MicroTymp3 portable tympanometric instrument
Open
The test has not begun since a valid ear cavity has not been detected.
Possible causes:
Instrument is not in an ear
Probe tip is not completely sealed in an ear
Perforated tympanic membrane
Ear with patent tympanostomy tubes
Extremely large canal (> 2.5 cc)
Possible solutions:
Reposition the probe tip
Perform otoscopy to check for perforated membrane or patent tympanostomy tubes
Test
The TEST message indicates that the test has started. Test data will begin to appear
immediately.